Suppr超能文献

Interrupted coalescent nitinol clip versus continuous suture coronary anastomosis: a comparative endothelial function study.

作者信息

Demaria Roland G, Fortier Simon, Malo Olivier, Carrier Michel, Perrault Louis P

机构信息

Department of Surgery and Research Center, Montreal Heart Institute, Montreal, Quebec, Canada.

出版信息

Heart Surg Forum. 2003;6(2):72-6. doi: 10.1532/hsf.1058.

Abstract

BACKGROUND

A new penetrating stapled anastomotic system using nitinol microclips (Coalescent Surgical U-Clip) has been developed to facilitate the construction of compliant interrupted sutures for minimally invasive and robotic surgery as well as for conventional procedures. The purpose of this study was to determine the effect of nitinol U-Clips used for coronary anastomosis in the development of endothelial dysfunction, compared with conventional running sutures.

METHODS

In a porcine model, both internal thoracic arteries were harvested, and the heart was removed. In a Krebs solution, 2 anastomoses were realized ex vivo between internal thoracic arteries and the left anterior descending artery. One was carried out with 12 Coalescent microclips, and the other used conventional running 7-0 polypropylene suture material (Prolene). Coronary rings on the anastomotic sites were then placed in organ chambers filled with oxygenated Krebs solution. Vascular reactivity studies were performed in standard organ chamber experiments. After the contraction of the coronary arteries in response to prostaglandin F2alpha, the endothelium-dependent relaxation response to bradykinin was studied. The other coronary arteries served as controls.

RESULTS

There was no statistically significant difference among the groups (P >.05) in the amplitude of the contraction response to KCl and prostaglandin F2alpha. There were no statistically significant differences in endothelium-dependent relaxation response to bradykinin between the nitinol microclip group and controls, between the suture group and controls, and between the nitinol microclip and suture groups.

CONCLUSIONS

Coalescent nitinol U-Clips used as anastomotic devices do not induce an endothelial dysfunction and allow a compliant anastomosis under satisfactory conditions.

摘要

相似文献

2
Nonpenetrating stapling: a valuable alternative for coronary anastomoses?
Ann Thorac Surg. 1998 Nov;66(5):1705-8. doi: 10.1016/s0003-4975(98)00912-6.
3
Interrupted nitinol U-Clips versus standard running suture for the central arterial T-graft anastomosis: a prospective randomized study.
Eur J Cardiothorac Surg. 2011 Aug;40(2):e93-7. doi: 10.1016/j.ejcts.2011.02.078. Epub 2011 May 6.
4
Facilitated coronary anastomosis using a nitinol U-Clip device: bovine model.
J Thorac Cardiovasc Surg. 2001 May;121(5):859-70. doi: 10.1067/mtc.2001.112936.
5
Endothelial effects of hemostatic devices for continuous cardioplegia or minimally invasive operations.
Ann Thorac Surg. 1996 Oct;62(4):1158-63. doi: 10.1016/0003-4975(96)00536-X.
7
A prospective evaluation of interrupted nitinol surgical clips in arteriovenous fistula for hemodialysis.
Am J Surg. 2003 Dec;186(6):625-30. doi: 10.1016/j.amjsurg.2003.08.007.
10
Snaring of the target vessel in less invasive bypass operations does not cause endothelial dysfunction.
Ann Thorac Surg. 1997 Mar;63(3):751-5. doi: 10.1016/s0003-4975(96)01118-6.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验