Imamura Keiko, Terada Hiroshi, Hagiwara Akira, Higashida Yoshiharu, Ehara Norishige, Ohuchi Noriaki
Department of Radiology, St. Marianna University School of Medicine.
Igaku Butsuri. 2000;20(1):46-55.
To quantitatively demonstrate inconsistencies in microcalcification specks in RMI-156 phantoms approved by the American College of Radiology (ACR) for use with their mammography accreditation program. MATERIALS AND METHODS: Signal intensities of the largest three specks (C1, 540 &mgr;m; C2, 400 &mgr;m; C3, 320 &mgr;m) were measured on images of 14 RMI-156 (version 4) phantoms. Three images were analyzed per phantom: a packaged image of the wax plate and two phantom images obtained at an optical density of 1.2 to 1.3. Phantom images were digitized using a VXR-12 scanner at a resolution of 85 &mgr;m and 8 bit/pixel. Speck signals were measured as the average pixel values using the NIH Image program (version 1.61). RESULTS: Signal intensities of specks measured on phantom images were consistent with those measured on wax-plate images, with an average correlation coefficient of 0.898 for the C1 groups and of 0.794 for the C3 groups. Overlap of speck sizes was found between different groups. Speck signals of some inserts showed considerable inconsistencies within a group. In some phantoms, signal intensities of specks in different groups overlapped to each other, and C1 and C2 groups were not clearly separated. In all 14 phantoms, speck signal intensities of the C3 group had a between-phantom variance which was larger than the within-phantom variance. CONCLUSION: Microcalcification specks in RMI-156 phantoms of the same version were found to be substantially inconsistent in terms of size and intrinsic signal intensity. As lack of uniformity in specks within a phantom and between phantoms is not appropriate as a standard of image quality control, further quality control is apparently needed in manufacturing RMI-156 phantoms and also in approving specific phantoms.
定量证明美国放射学会(ACR)批准用于其乳腺摄影认证计划的RMI - 156体模中微钙化斑点的不一致性。材料与方法:在14个RMI - 156(版本4)体模的图像上测量最大的三个斑点(C1,540μm;C2,400μm;C3,320μm)的信号强度。每个体模分析三张图像:蜡板的包装图像以及在光密度为1.2至1.3时获得的两张体模图像。使用VXR - 12扫描仪以85μm的分辨率和8位/像素对体模图像进行数字化处理。使用NIH Image程序(版本1.61)将斑点信号测量为平均像素值。结果:在体模图像上测量的斑点信号强度与在蜡板图像上测量的信号强度一致,C1组的平均相关系数为0.898,C3组的平均相关系数为0.794。发现不同组之间存在斑点大小的重叠。一些插入物的斑点信号在组内显示出相当大的不一致性。在一些体模中,不同组的斑点信号强度相互重叠,并且C1和C2组没有明显分开。在所有14个体模中,C3组的斑点信号强度在体模间的方差大于体模内的方差。结论:发现同一版本的RMI - 156体模中的微钙化斑点在大小和固有信号强度方面存在很大不一致性。由于体模内和体模间斑点缺乏均匀性不适宜作为图像质量控制标准,显然在制造RMI - 156体模以及批准特定体模时需要进一步的质量控制。