一种胸部体模的开发与应用,用于优化多种低剂量螺旋计算机断层扫描设备的扫描技术。

The development and use of a chest phantom for optimizing scanning techniques on a variety of low-dose helical computed tomography devices.

作者信息

Muramatsu Yoshihisa, Tsuda Yukihiro, Nakamura Yoshimasa, Kubo Mitsuru, Takayama Toshiyuki, Hanai Kouzou

机构信息

Department of Dignostic Radiology, national Cancer Center Hospital, Tokyo, Japan.

出版信息

J Comput Assist Tomogr. 2003 May-Jun;27(3):364-74. doi: 10.1097/00004728-200305000-00012.

Abstract

PURPOSE

To develop a chest phantom for determining the optimal scan conditions for chest computed tomography (CT) screening.

METHODS

The basic structure of the phantom is an arms-elevated-positioned anthropomorphic chest phantom. The internal structure includes simulated tumors (which are assumed to be the target lesions of chest CT screening examinations) placed at the levels of the lung apex, the bifurcation of the trachea, and the lung base of both simulated lungs. The image contrast of the simulated tumors is taken as the difference in CT value and is specified as 1 of 2 target values: Delta270 and Delta100. An opening for placement of a dosimeter is provided on the central axis of the phantom. Initial field tests were conducted focusing on the scan conditions for chest CT screening.

RESULTS

Images similar to those obtained in chest CT screening examinations and clearly showing pathologic changes were obtained. The dose measurement at the center was 2.0 mGy. The diameters of the simulated tumors that could be detected were 6 mm for Delta270 and 10 mm for Delta100.

CONCLUSION

The use of this phantom makes it possible to determine the optimal scan conditions for chest CT screening based on objective evaluation criteria.

摘要

目的

开发一种胸部体模,用于确定胸部计算机断层扫描(CT)筛查的最佳扫描条件。

方法

体模的基本结构是一个双臂上举姿势的仿真人体胸部体模。内部结构包括放置在双侧模拟肺的肺尖、气管分叉处和肺底部水平的模拟肿瘤(假定为胸部CT筛查检查的目标病变)。模拟肿瘤的图像对比度以CT值之差表示,并指定为2个目标值之一:Delta270和Delta100。在体模的中心轴上设有一个用于放置剂量计的开口。针对胸部CT筛查的扫描条件进行了初步现场测试。

结果

获得了与胸部CT筛查检查中获得的图像相似且能清晰显示病理变化的图像。中心部位的剂量测量值为2.0 mGy。对于Delta270,可检测到的模拟肿瘤直径为6 mm;对于Delta100,可检测到的模拟肿瘤直径为10 mm。

结论

使用这种体模能够基于客观评估标准确定胸部CT筛查的最佳扫描条件。