罗氏LightCycler的评估:一种无需提取DNA即可直接从血样中检测因子V莱顿突变和凝血酶原G20210A基因型的简单快速方法。

Evaluation of the Roche LightCycler: a simple and rapid method for direct detection of factor V Leiden and prothrombin G20210A genotypes from blood samples without the need for DNA extraction.

作者信息

Cooper Peter C, Cooper Susan M, Smith Julie M, Kitchen Steven, Makris Michael

机构信息

Sheffield Haemophilia and Thrombosis Centre, Royal Hallamshire Hospital, UK.

出版信息

Blood Coagul Fibrinolysis. 2003 Jul;14(5):499-503. doi: 10.1097/01.mbc.0000061327.06975.ae.

Abstract

The Roche LightCycler is a micro-volume thermocycler that combines extremely rapid polymerase chain reaction with fluorescence resonance energy transfer analysis of amplified products. We have evaluated the use of minimally processed blood samples for detection of two point mutations known to increase the risk of venous thromboembolism. Results from the LightCycler using the supernatant of diluted heated blood were compared with those gained by traditional methods based on polymerase chain reaction and restriction enzyme digestion. For factor V Leiden mutation, there was complete agreement between both methods in detection of wild-type (n = 82), heterozygous (n = 100) and homozygous (n = 18) genotypes. Similarly, the prothrombin G20210A mutation showed complete agreement for wild-type (n = 135), heterozygous (n = 63) and homozygous (n = 2) subjects.

摘要

罗氏LightCycler是一种微量热循环仪,它将极快速的聚合酶链反应与扩增产物的荧光共振能量转移分析相结合。我们评估了使用最少处理的血样来检测已知会增加静脉血栓栓塞风险的两种点突变。将使用稀释加热血样上清液的LightCycler检测结果与基于聚合酶链反应和限制性酶切的传统方法所获结果进行比较。对于因子V莱顿突变,两种方法在检测野生型(n = 82)、杂合子(n = 100)和纯合子(n = 18)基因型方面完全一致。同样,凝血酶原G20210A突变在野生型(n = 135)、杂合子(n = 63)和纯合子(n = 2)受试者中也显示出完全一致。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验