Suppr超能文献

用于验证多能量混合调强放疗治疗计划的归一化剂量测定曲线。

Normalized sensitometric curves for the verification of hybrid IMRT treatment plans with multiple energies.

作者信息

Georg Dietmar, Kroupa Bernhard, Winkler Peter, Pötter Richard

机构信息

Division of Medical Radiation Physics, Department of Radiotherapy and Radiobiology, University of Vienna Medical School, Waehringer Guertel 18-20, A-1090 Vienna, Austria.

出版信息

Med Phys. 2003 Jun;30(6):1142-50. doi: 10.1118/1.1576951.

Abstract

With the clinical implementation of time-variable dose patterns and intensity modulated radiotherapy (IMRT) film dosimetry has regained popularity. Films are currently the most frequently used dosimetric means for patient specific quality assurance in IMRT. A common method is to verify a so-called hybrid IMRT plan, which is the patient specific treatment plan with unmodified fluence patterns recalculated in a dedicated phantom. For such applications the sensitometric curve, i.e., the relation between optical density (OD) and absorbed dose, should not depend critically on beam energy, field size and depth, or film orientation. In order to minimize the influence of all these variables a normalization of sensitometric curves is performed at various photon beam energies (6 MV, 10 MV, 25 MV). By doing so one unique sensitometric curve can be used for these three beam qualities. This holds for both film types investigated: Kodak X-Omat V films and EDR-2 films. Additionally, the influence of field size, depth and film orientation on a normalized sensitometric curve is determined for both film types. For doses smaller than 0.8 Gy for X-Omat V and doses smaller than 3 Gy for EDR-2 films the field size variation of normalized sensitometric curves is much smaller than 3% for fields up to 20 x 20 cm2. For X-Omat V films all differences between sensitometric curves determined at depths of 5, 10, and 15 cm are smaller than 3%. For EDR-2 films deviations larger than 3% are only observed at low net OD smaller than 0.25. The dependence of film orientation (parallel versus perpendicular) on a normalized sensitometric curve is found to be not critical. However, processing conditions have the largest influence and can result in differences up to 20% for sensitometric curves derived from films of the same batch but using different film processors. When normalizing sensitometric curves to the dose value necessary to obtain a net OD=1 for that respective geometry and energy the large energy dependence of sensitometric curves can be almost eliminated. This becomes especially important for the verification of hybrid IMRT plans with multiple energies. Additionally, such a normalization minimizes other influences such as field size, depth, and film orientation. This method is generally applicable to both Kodak X-Omat V and EDR-2 films. In order to achieve the highest accuracy level an upper dose limit of 0.8 Gy for X-Omat V films and 3 Gy for EDR-2 films should be taken into account. However, these dose limits may vary with film reading instrument and film processor.

摘要

随着时变剂量模式和调强放疗(IMRT)在临床上的应用,胶片剂量测定法再度受到关注。目前,胶片是IMRT中最常用于患者特定质量保证的剂量测定手段。一种常用方法是验证所谓的混合IMRT计划,即在专用模体中重新计算通量模式未修改的患者特定治疗计划。对于此类应用,感光曲线,即光密度(OD)与吸收剂量之间的关系,不应严重依赖于束能量、射野大小和深度或胶片方向。为了最小化所有这些变量的影响,在不同光子束能量(6 MV、10 MV、25 MV)下对感光曲线进行归一化。这样,一条独特的感光曲线可用于这三种束质量。这适用于所研究的两种胶片类型:柯达X-Omat V胶片和EDR-2胶片。此外,还确定了两种胶片类型的射野大小、深度和胶片方向对归一化感光曲线的影响。对于X-Omat V胶片,剂量小于0.8 Gy,对于EDR-2胶片,剂量小于3 Gy时,对于面积达20×20 cm2的射野,归一化感光曲线的射野大小变化远小于3%。对于X-Omat V胶片,在5 cm、10 cm和15 cm深度处测定的感光曲线之间的所有差异均小于3%。对于EDR-2胶片,仅在净OD小于0.25的低剂量时观察到大于3%的偏差。发现胶片方向(平行与垂直)对归一化感光曲线的依赖性不显著。然而,处理条件的影响最大,对于同一批次但使用不同胶片处理器的胶片得出的感光曲线,可能导致高达20%的差异。当将感光曲线归一化为在该特定几何形状和能量下获得净OD = 1所需的剂量值时,感光曲线的大能量依赖性几乎可以消除。这对于验证具有多种能量的混合IMRT计划尤为重要。此外,这种归一化最小化了诸如射野大小、深度和胶片方向等其他影响。该方法通常适用于柯达X-Omat V和EDR-2胶片。为了达到最高精度水平,应考虑X-Omat V胶片的剂量上限为0.8 Gy,EDR-2胶片的剂量上限为3 Gy。然而,这些剂量限制可能因胶片读取仪器和胶片处理器而异。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验