Berry Julianne, Kline Lukeysha C, Hart John L, Sequeira Joel
Schering Plough Research Institute, Kenilworth, New Jersey 07033, USA.
Drug Dev Ind Pharm. 2003 Jul;29(6):631-9. doi: 10.1081/ddc-120021312.
Presented in this work are the results of a study designed to investigate the impact of the storage position on the particle size distribution (PSD) of a steroid suspension metered dose inhaler (MDI) containing propellant HFA-227. It was hypothesized that the orientation of MDI samples upon storage could influence the PSD of the emitted dose, since it determines the amount of contact the liquid formulation has with the valve and therefore the quantity of nonvolatile leachable materials from the valve components that may enter the product and potentially impact the aerosol spray dynamics. Samples stored in the valve down orientation (i.e., complete contact of the liquid formulation with the valve) showed a higher level of leachables compared to those samples stored valve up (i.e., minimal contact of the formulation with the valve). The valve down samples were found to produce larger particles in the emitted aerosol spray using both cascade impaction, the preferred method of regulatory submission, as well as laser diffraction. It was postulated that the larger particle size of the inverted samples was attributed to its higher levels of leachables. Based on our findings, it is recommended that in order to set appropriate controls on the product PSD, the storage orientation of the product will need to be considered.
本研究旨在调查储存位置对含有推进剂HFA - 227的类固醇混悬液定量吸入器(MDI)的粒度分布(PSD)的影响,本文展示了该研究的结果。研究假设,MDI样品储存时的方向可能会影响喷出剂量的PSD,因为它决定了液体制剂与阀门的接触量,进而决定了可能进入产品并潜在影响气溶胶喷雾动力学的阀门部件中非挥发性可浸出材料的量。与阀门向上储存的样品(即制剂与阀门的接触最少)相比,阀门向下储存的样品(即液体制剂与阀门完全接触)显示出更高水平的可浸出物。使用级联冲击法(监管申报的首选方法)以及激光衍射法发现,阀门向下的样品在喷出的气溶胶喷雾中产生的颗粒更大。据推测,倒置样品的较大粒径归因于其较高水平的可浸出物。基于我们的研究结果,建议为了对产品的PSD进行适当控制,需要考虑产品的储存方向。