Krajcar Zvonimir, Gupta Kamal, Dougherty Kathy G
Department of Cardiology, St. Luke' Episcopal Hospital and Texas Heart Institute, Houston, Texas 77030, USA.
Tex Heart Inst J. 2003;30(3):186-93.
We present a series of 4 patients in whom mechanical trauma was identified as a factor in the development of late complications after AneuRx Stent Graft placement for repair of abdominal aortic aneurysms. In all 4 patients, Type I or III endoleaks (and pseudoaneurysms in 2 patients) were discovered several months after abdominal aortic aneurysm repair with the AneuRx device. Two patients had sustained blunt abdominal trauma in a car accident one had suffered a traumatic fall, and another had been participating in vigorous rowing activity. In all patients, the trauma had occurred several months before the diagnosis of endoleak or pseudoaneurysm (or both) was established. In all patients, follow-up computed tomographic scans identified the complications. In conclusion, blunt mechanical injury is an unrecognized factor contributing to the late failure of endovascular stent grafts. Vigorous physical activity may also contribute to graft disruption or to the separation of modular components.
我们报告了4例患者,在这些患者中,机械性创伤被确定为使用AneuRx覆膜支架修复腹主动脉瘤后发生晚期并发症的一个因素。在所有4例患者中,使用AneuRx装置修复腹主动脉瘤数月后发现了I型或III型内漏(2例患者还伴有假性动脉瘤)。2例患者在车祸中遭受腹部钝性创伤,1例患者有外伤性跌倒,另1例患者一直在进行剧烈的划船活动。在所有患者中,创伤发生在诊断内漏或假性动脉瘤(或两者)之前数月。在所有患者中,随访计算机断层扫描确定了并发症。总之,钝性机械损伤是导致血管内覆膜支架晚期失效的一个未被认识的因素。剧烈的体力活动也可能导致移植物破裂或模块化组件分离。