Computer-assisted two-dimensional analysis of gastric mucosal hemoglobin distribution using electronic endoscopy.

作者信息

Sato N, Tsuji S, Kawano S, Hayashi N, Tsujii M, Ishigami Y, Ogihara T, Nagano K, Kamada T

机构信息

Department of Gastroenterology, Juntendo University School of Medicine, Tokyo, Japan.

出版信息

Endoscopy. 1992 Jul;24 Suppl 2:522-6. doi: 10.1055/s-2007-1010535.

Abstract

An electronic endoscope was modified to assess the imaging of two-dimensional distribution of gastrointestinal mucosal hemoglobin with the aid of a computer and image analyzing system. The band-pass filters ranging from 530-470 nm (G'), and 770-830 nm (IR) replaced ordinary filters in a conventional electronic endoscope (Olympus V-10 or EVIS-200). As the hemoglobin is the most abundant biological pigment in the gastrointestinal mucosa absorbing the green light with little absorption in the infrared region, we used the algorithm of 32 [log2(Vir/Vg)] as a reflection of hemoglobin content in the surface mucosa. The algorithm had a good correlation with the spectrophotometrically measured hemoglobin concentration. The mucosal distribution of hemoglobin was displayed in patients with gastric ulcer. In conclusion, the computer and image analyzing system were successfully applied in an attempt to determine gastric mucosal hemoglobin content and its distribution. The possible applications of computers in endoscopic image analysis are discussed.