Bhuta Ujvala M, Ulstein Howard
Florida Medical Center, Fort Lauderdale, Florida 33313, USA.
Lab Hematol. 2003;9(3):167-74.
The Coulter AcT 5-part differential (5 diff) autoloader (AL) hematology analyzer from Beckman Coulter (Fullerton, CA, USA) was evaluated at the Florida Medical Center in Fort Lauderdale, Florida, an acute care hospital facility. The AcT 5 diff AL is a new, fully automated bench-top 5-part differential hematology analyzer with automatic loading and sampling. It is designed as a front-line instrument for small- to medium-sized laboratories or as a backup in larger laboratories. We evaluated the performance of the AcT 5 diff AL for complete blood counts (CBCs). The Coulter Onyx AL was used as the reference instrument and the 5-part differential against the reference 400-cell manual differential count (2 operators x 200 cells). A total of 140 patient samples were analyzed in the study, with most of the samples obtained from our own laboratory workload. We supplemented this dataset with pediatric and oncology samples from nearby practices. All samples were analyzed within 4 hours of collection, and the AcT 5 diff AL demonstrated excellent accuracy with correlation values (r) > 0.98 for all directly measured CBC parameters (white blood cell count, red blood cell count, hemoglobin, mean corpuscular volume, and platelets). Reproducibility studies on the AcT 5 diff AL also showed excellent results, with coefficients of variation much lower than the manufacturer's specifications. Initial differential comparisons showed good correlations for neutrophils, lymphocytes, and eosinophils (r = 0.84, 0.96, and 0.88, respectively) with a lower correlation for monocytes (r = 0.43). On subsequent review and data analysis we found differential discrepancies in 39 of the 140 samples, for which all showed multiple morphological flags, R diff flags, or poor scatterplot separations indicating that the AcT 5 diff AL differential was likely to be inaccurate and that a morphological assessment and differential was necessary. Given the acutecare setting that our laboratory serves and the skewing of the dataset of our study because of the addition of abnormal pediatric and oncology cases, these abnormal sample results were not surprising. These 39 samples all had markedly abnormal differentials with either increased numbers of immature and/or abnormal cells due to such conditions as end-stage liver disease, respiratory failure, renal failure, sepsis and pneumonia, high white blood cell leukemias, and a variety of other acute illnesses. Reanalysis of the differential data, excluding these 39 grossly abnormal samples with R flags and multiple flags, showed the AcT 5 diff AL differential gave very good correlations for neutrophils, lymphocytes, monocytes, and eosinophils (r = 0.93, 0.97, 0.66, and 0.85, respectively). Our technologists were also impressed with the reliability of the AcT 5 diff AL, its ease of use and transfer from stand-by to ready mode, the user friendliness of the software, and the clarity of the printed report. We concluded that the AcT 5 diff AL gives accurate and precise CBC and 5-part differential results and is an excellent, easy-to-use instrument ideally suited for small- to moderate-sized hospital laboratories, as a back-up in a large hospital laboratory, or for use in physician office laboratories.
对来自美国加利福尼亚州富勒顿市贝克曼库尔特公司的库尔特AcT 5分类(5 diff)自动进样器(AL)血液分析仪进行了评估,评估地点为佛罗里达州劳德代尔堡的佛罗里达医疗中心,这是一家急症护理医院机构。AcT 5 diff AL是一款新型的、全自动的台式5分类血液分析仪,具备自动进样和取样功能。它被设计用作中小型实验室的一线仪器,或大型实验室的备用仪器。我们评估了AcT 5 diff AL进行全血细胞计数(CBC)的性能。将库尔特玛瑙AL用作参考仪器,并将5分类结果与参考的400个细胞的手工分类计数(2名操作人员×200个细胞)进行对比。本研究共分析了140份患者样本,其中大部分样本取自我们自己实验室的工作量。我们用附近医疗机构的儿科和肿瘤样本对该数据集进行了补充。所有样本均在采集后4小时内进行分析,AcT 5 diff AL对所有直接测量的CBC参数(白细胞计数、红细胞计数、血红蛋白、平均红细胞体积和血小板)的相关性值(r)>0.98,显示出极佳的准确性。对AcT 5 diff AL的重复性研究也显示出优异的结果,变异系数远低于制造商的规格。初始分类比较显示,中性粒细胞、淋巴细胞和嗜酸性粒细胞的相关性良好(r分别为0.84、0.96和0.88),单核细胞的相关性较低(r = 0.43)。在后续的复查和数据分析中,我们在140份样本中的39份中发现了分类差异,所有这些样本均显示出多个形态学标记、R diff标记或散点图分离不佳,表明AcT 5 diff AL的分类可能不准确,需要进行形态学评估和分类。考虑到我们实验室服务的急症护理环境,以及由于添加了异常儿科和肿瘤病例而导致我们研究数据集的偏差,这些异常样本结果并不令人惊讶。这39份样本的分类均明显异常,由于终末期肝病、呼吸衰竭、肾衰竭、败血症和肺炎、高白细胞白血病以及各种其他急性疾病等情况,未成熟和/或异常细胞数量增加。排除这39份带有R标记和多个标记的严重异常样本后,对分类数据进行重新分析,结果显示AcT 5 diff AL对中性粒细胞、淋巴细胞、单核细胞和嗜酸性粒细胞的分类相关性非常好(r分别为0.93、0.97、0.66和0.85)。我们的技术人员也对AcT 5 diff AL的可靠性、从待机模式转换到就绪模式的便捷性、软件的用户友好性以及打印报告的清晰度印象深刻。我们得出结论,AcT 5 diff AL能给出准确、精确的CBC和5分类结果,是一款出色的、易于使用的仪器,非常适合中小型医院实验室、大型医院实验室的备用仪器或用于医生办公室实验室。