Suppr超能文献

非充盈状态下人体左心室的舒张功能。

Diastolic function of the nonfilling human left ventricle.

作者信息

Paulus W J, Vantrimpont P J, Rousseau M F

机构信息

Cardiovascular Center, O.L.V. Ziekenhuis, Aalst, Belgium.

出版信息

J Am Coll Cardiol. 1992 Dec;20(7):1524-32. doi: 10.1016/0735-1097(92)90446-t.

Abstract

OBJECTIVES

To investigate an early-diastolic left ventricular suction effect in humans, tip-micromanometer left ventricular pressure recordings were obtained in patients with mitral stenosis at the time of balloon inflations during percutaneous mitral valvuloplasty performed with a self-positioning Inoue balloon, which fits tightly in the mitral orifice.

BACKGROUND

When mitral inflow was impeded in anesthetized dogs, left ventricular pressure decayed to a negative asymptote value. This negative asymptote value was consistent with an early diastolic suction effect.

METHODS

Tip-micromanometer left ventricular pressure recordings were obtained in 23 patients with symptomatic mitral stenosis at the time of balloon inflations during percutaneous mitral valvuloplasty performed with a self-positioning Inoue balloon.

RESULTS

The left ventricular diastolic asymptote pressure (P(asy)) was determined in 47 nonfilling beats with a sufficiently long (greater than 200 ms) diastolic time interval (that is, the interval from minimal first derivative of left ventricular pressure to left ventricular end-diastolic pressure) and equaled 2 +/- 3 mm Hg for beats with normal intraventricular conduction and 3 +/- 2 mm Hg for beats with aberrant intraventricular conduction. Left ventricular angiography was performed in five patients during the first inflation of the Inoue balloon at the time of complete balloon expansion. Left ventricular end-diastolic volume of the nonfilling beats averaged 38 +/- 14 ml and was comparable to the left ventricular end-systolic volume (39 +/- 19 ml) measured during baseline angiography before mitral valvuloplasty. Time constants of left ventricular pressure decay were calculated on 21 nonfilling beats with a diastolic time interval greater than 200 ms, normal intraventricular conduction and peak left ventricular pressure greater than 50 mm Hg. Time constants (T0 and TBF) derived from an exponential curve fit with zero asymptote pressure and with a best-fit asymptote pressure were compared with a time constant (T(asy)) derived from an exponential curve fit with the measured diastolic left ventricular asymptote pressure. The value for T(asy) (37 +/- 9 ms) was significantly smaller than that for TBF (68 +/- 28 ms, p less than 0.001) and the value for the measured diastolic left ventricular asymptote pressure (2 +/- 4 mm Hg) was significantly larger than that for the best-fit asymptote pressure (-9 +/- 11 mm Hg, p less than 0.001). T0 (44 +/- 20 ms) was significantly (p less than 0.01) different from TBF but not from T(asy).

CONCLUSIONS

During balloon inflation of a self-positioning Inoue balloon, left ventricular pressure decayed continuously toward a positive asymptote value and left ventricular cavity volume was comparable to the left ventricular end-systolic volume of filling beats. In these nonfilling beats, the best-fit asymptote pressure was unrelated to the measured asymptote pressure and T0 was a better measure of T(asy) than was TBF. Reduced internal myocardial restoring forces, caused by different extracellular matrix of the human heart, reduced external myocardial restoring forces caused by low coronary perfusion pressure during the balloon inflation and inward motion of the balloon-occluded mitral valve into the left ventricular cavity could explain the failure to observe significant diastolic left ventricular suction in the human heart.

摘要

目的

为研究人体舒张早期左心室抽吸效应,在使用能紧密贴合二尖瓣口的自定位井上球囊进行经皮二尖瓣成形术时,对二尖瓣狭窄患者球囊充盈时进行尖端微测压左心室压力记录。

背景

在麻醉犬中,当二尖瓣血流受阻时,左心室压力衰减至负渐近值。该负渐近值与舒张早期抽吸效应一致。

方法

在23例有症状的二尖瓣狭窄患者使用自定位井上球囊进行经皮二尖瓣成形术时,于球囊充盈时进行尖端微测压左心室压力记录。

结果

在47次非充盈搏动中确定了左心室舒张渐近压力(P(asy)),这些搏动具有足够长(大于200毫秒)的舒张时间间隔(即从左心室压力的最小一阶导数到左心室舒张末期压力的间隔),对于室内传导正常的搏动,P(asy)等于2±3毫米汞柱,对于室内传导异常的搏动,P(asy)等于3±2毫米汞柱。在井上球囊首次充盈且完全扩张时,对5例患者进行了左心室造影。非充盈搏动的左心室舒张末期容积平均为38±14毫升,与二尖瓣成形术前基线造影时测得的左心室收缩末期容积(39±19毫升)相当。在21次舒张时间间隔大于200毫秒、室内传导正常且左心室峰值压力大于50毫米汞柱的非充盈搏动上计算左心室压力衰减的时间常数。将从零渐近压力和最佳拟合渐近压力的指数曲线拟合得出的时间常数(T0和TBF)与从测量的舒张期左心室渐近压力的指数曲线拟合得出的时间常数(T(asy))进行比较。T(asy)的值(37±9毫秒)显著小于TBF的值(68±28毫秒,p<0.001),且测量的舒张期左心室渐近压力的值(2±4毫米汞柱)显著大于最佳拟合渐近压力的值(-9±11毫米汞柱,p<0.001)。T0(44±20毫秒)与TBF有显著差异(p<0.01),但与T(asy)无显著差异。

结论

在自定位井上球囊充盈过程中,左心室压力持续衰减至正渐近值,且左心室腔容积与充盈搏动的左心室收缩末期容积相当。在这些非充盈搏动中,最佳拟合渐近压力与测量的渐近压力无关,且T0比TBF更能准确测量T(asy)。人体心脏不同的细胞外基质导致心肌内部恢复力降低、球囊充盈时低冠状动脉灌注压力导致心肌外部恢复力降低以及球囊阻塞的二尖瓣向左心室腔内的向内运动,可能解释了在人体心脏中未观察到显著舒张期左心室抽吸的原因。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验