Bellieni Carlo V, Buonocore Giuseppe, Pinto Iole, Stacchini Nicola, Cordelli Duccio M, Bagnoli Franco
Department of Pediatrics, Obstetrics and Reproduction Medicine, University of Siena, Siena, Italy.
Biol Neonate. 2003;84(4):293-6. doi: 10.1159/000073637.
To assess the amplification and distortion of noise in incubators by measuring noise levels from various sources under various conditions.
We recorded sound pressure levels (SPL) of background noise, baby crying, alarms and porthole closing with the incubator hood open or closed. Then we tried to reduce the sound level in the incubator by applying a sound absorbing panel to the ceiling of the hood.
The sound sources analysed produced very high SPL. In free field baby crying and porthole noise were, respectively, 81-83 dB and 70-71 dB, closing the hood they raised to 84-87 and 73-74 dB: this means that the noise perceived by a baby in the incubator had twice the SPL of the same noise emitted in a free field; alarm noise was 56-57 dB inside incubators. Sound emitted inside the hood was also distorted, i.e. it had different features from the same source in a free field. The insulating panel prevented amplification and distortion due to the hood and reduced the SPL of alarms. Only background noise did not decrease with the panel. Sound absorbing panel reduced baby's crying, porthole noise and alarm noise SPL to 82-85, 70-71 and 50-51 dB, respectively.
Incubators are reverberating environments, which amplify the baby's cry and other noises produced inside the hood. The frequency components of noises are altered by the hood. The sound absorbing panel reduced this phenomenon.
通过测量各种条件下不同声源的噪声水平,评估培养箱内噪声的放大和失真情况。
我们记录了培养箱罩打开或关闭时背景噪声、婴儿哭声、警报声和舱门关闭声的声压级(SPL)。然后,我们试图通过在培养箱罩的天花板上安装吸音板来降低培养箱内的声音水平。
分析的声源产生了非常高的声压级。在自由场中,婴儿哭声和舱门噪声分别为81 - 83分贝和70 - 71分贝,关闭培养箱罩后,它们分别升至84 - 87分贝和73 - 74分贝:这意味着培养箱内婴儿所感知到的噪声声压级是在自由场中发出的相同噪声的两倍;培养箱内警报噪声为56 - 57分贝。培养箱罩内发出的声音也发生了失真,即与自由场中同一声源的声音具有不同的特征。绝缘板可防止因培养箱罩导致的放大和失真,并降低警报声的声压级。只有背景噪声不会因该面板而降低。吸音板将婴儿哭声、舱门噪声和警报噪声的声压级分别降低到82 - 85分贝,70 - 71分贝和50 - 51分贝。
培养箱是混响环境,会放大婴儿哭声和培养箱罩内产生的其他噪声。噪声的频率成分会因培养箱罩而改变。吸音板减少了这种现象。