Accurate wavelength measurements of a putative standard for near-infrared diffuse reflection spectrometry.

作者信息

Isaksson Tomas, Yang Husheng, Kemeny Gabor J, Jackson Richard S, Wang Qian, Alam M Kathleen, Griffiths Peter R

机构信息

Department of Chemistry, University of Idaho, Moscow, Idaho 83844-2343, USA.

出版信息

Appl Spectrosc. 2003 Feb;57(2):176-85. doi: 10.1366/000370203321535097.

Abstract

The diffuse reflection (DR) spectrum of a sample consisting of a mixture of rare earth oxides and talc was measured at 2 cm-1 resolution, using five different accessories installed on five different Fourier transform near-infrared (FT-NIR) spectrometers from four manufacturers. Peak positions for 37 peaks were determined using two peak-picking algorithms: center-of-mass and polynomial fitting. The wavenumber of the band center reported by either of these techniques was sensitive to the slope of the baseline, and so the baseline of the spectra was corrected using either a polynomial fit or conversion to the second derivative. Significantly different results were obtained with one combination of spectrometer and accessory than the others. Apparently, the beam path through the interferometer and DR accessory was different for this accessory than for any of the other measurements, causing a severe degradation of the resolution. Spectra measured on this instrument were removed as outliers. For measurements made on FT-NIR spectrometers, it is shown that it is important to check the resolution at which the spectrum has been measured using lines in the vibration-rotation spectrum of atmospheric water vapor and to specify the peak-picking and baseline-correction algorithms that are used to process the measured spectra. The variance between the results given by the four different methods of peak-picking and baseline correction was substantially larger than the variance between the remaining five measurements. Certain bands were found to be more suitable than others for use as wavelength standards. A band at 5943.13 cm-1 (1682.62 nm) was found to be the most stable band between the four methods and the six measurements. A band at 5177.04 cm-1 (1931.61 nm) has the highest precision between different measurements when polynomial baseline correction and polynomial peak-picking algorithms are used.

摘要

使用安装在来自四个制造商的五台不同傅里叶变换近红外(FT-NIR)光谱仪上的五种不同附件,以2 cm⁻¹ 的分辨率测量了由稀土氧化物和滑石粉混合物组成的样品的漫反射(DR)光谱。使用两种峰提取算法(质心算法和多项式拟合)确定了37个峰的峰位置。这两种技术报告的谱带中心波数对基线斜率敏感,因此使用多项式拟合或转换为二阶导数对光谱基线进行校正。一种光谱仪和附件的组合所获得的结果与其他组合有显著差异。显然,该附件通过干涉仪和DR附件的光路与其他任何测量的光路不同,导致分辨率严重下降。在该仪器上测量的光谱作为异常值被剔除。对于在FT-NIR光谱仪上进行的测量,结果表明,使用大气水汽振动-转动光谱中的谱线检查测量光谱的分辨率,并指定用于处理测量光谱的峰提取和基线校正算法非常重要。四种不同的峰提取和基线校正方法给出的结果之间的方差明显大于其余五次测量之间的方差。发现某些谱带比其他谱带更适合用作波长标准。在四种方法和六次测量中,5943.13 cm⁻¹(1682.62 nm)处的谱带是最稳定的谱带。当使用多项式基线校正和多项式峰提取算法时,5177.04 cm⁻¹(1931.61 nm)处的谱带在不同测量之间具有最高的精度。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验