Hong De-fei, Peng Shu-you, Tong Li-min, Li Song-ying, Cai Xiu-jun
Department of General Surgery, Sir Run Run Shawl Hospital, Zhejiang University, Hangzhou 310016, China.
Zhonghua Wai Ke Za Zhi. 2003 Nov;41(11):849-51.
To seek a safe, efficient, and cost-effective technique for local thermo-ablation of hepatic cancer.
The livers from 16 healthy rabbits were thermocoagulated by diode-laser with scanner fiber tip, 6 w for 10 mins. At the same time, the temperatures were measured at 0, 5 and 10 mm from laser tip. The pre-thermocoagulative liver function was compared with that of 7 days post-thermocoagulation. The pathologic changes were also observed 1 month after laser thermocoagulation.
All the rabbits survived and hepatic tissue temperatures at 0, 5, 10 mm from laser tip reached 96.39 degrees C +/- 3.97 degrees C, 60.79 degrees C +/- 6.21 degrees C, 46.10 degrees C +/- 4.58 degrees C respectively after 10 minutes of thermocoagulation. There were no significant differences in liver function parameters between rabbits of pre-laser thermocoagulation and of post-laser thermocoagulation. Thermocoagulated necrosis of liver tissue with surrounding fibrosis in a diameter of 26.0 mm was formed. Light microscopy revealed coagulative necrosis in the center of the coagulated area without surviving hepatic cells.
The hepatic tissue can be coagulated safely and effectively by diode-laser with scanner fibertip, and such a technique may provide a new method for the treatment of hepatic carcinoma.
寻找一种安全、高效且经济有效的肝癌局部热消融技术。
用带扫描光纤头的二极管激光对16只健康兔肝脏进行热凝,功率6瓦,持续10分钟。同时,在距激光头0、5和10毫米处测量温度。比较热凝前与热凝后7天的肝功能。在激光热凝1个月后观察病理变化。
所有兔子均存活,热凝10分钟后,距激光头0、5、10毫米处的肝组织温度分别达到96.39℃±3.97℃、60.79℃±6.21℃、46.10℃±4.58℃。激光热凝前与热凝后兔子的肝功能参数无显著差异。形成了直径为26.0毫米的肝组织热凝坏死伴周围纤维化。光镜下可见凝固区中心为凝固性坏死,无存活肝细胞。
带扫描光纤头的二极管激光可安全有效地凝固肝组织,该技术可能为肝癌治疗提供一种新方法。