Suppr超能文献

皮下监测器心电图分析。

Analysis of electrocardiograms for subcutaneous monitors.

作者信息

Burke Martin C, Song Zhendong, Jenkins Janice, Alberts Michael, Del Priore Jean, Arzbaecher Robert

机构信息

Department of Medicine/Cardiology, Loyola University Medical Center, Maywood, IL 60153, USA.

出版信息

J Electrocardiol. 2003;36 Suppl:227-32. doi: 10.1016/j.jelectrocard.2003.09.064.

Abstract

UNLABELLED

We are developing a subcutaneous cardiac arrest monitor and alarm with electrodes that have spacing of a few centimeters. We hypothesize that closely spaced bipolar electrodes that provide QRS amplitudes of a millivolt or more in sinus rhythm (SR) will not provide similar amplitudes during ventricular tachyarrhythmia (VT/VF), and that an orthogonal set of electrode pairs in diagonal is necessary to produce signals of sufficient amplitude in order to distinguish these rhythms.

METHODS

Forty patients were studied during ICD implantation in the clinical electrophysiology laboratory. A square array of 9 electrodes with 4- to 10-cm spacing between adjacent electrodes was placed on the patient's anterior chest over the left heart. The center electrode in the left most column is located in the 5th intercostal space at the left sternal border. Eight-channel recordings were made in SR and during induced VT/VF from the 8 peripheral electrodes referenced to the central member of the array. From these recordings all 20 bipolar ECGs of adjacent (including diagonally adjacent) electrodes were constructed algebraically. QRS peak-peak amplitudes in SR and VT/VF were measured in each bipolar lead and in the spatial vector formed by summing the squares of each adjacent and orthogonal pair of these leads.

RESULTS

Electrode pairs that yielded optimal ECG amplitudes in SR were not always the pairs that yielded optimal amplitudes in VT/VF. But in every patient two orthogonal pairs could be found whose QRS vector amplitude is sufficient in both SR and VT/VF to separate the rhythms.

CONCLUSIONS

A patient-defined set of bipolar electrode pairs is suitable for automatic separation of SR and VT/VF by rate in subcutaneous ambulatory monitoring.

摘要

未标注

我们正在研发一种皮下心脏骤停监测器及警报器,其电极间距为几厘米。我们推测,在窦性心律(SR)中能提供毫伏或更高QRS波幅的紧密间隔双极电极,在室性快速心律失常(VT/VF)期间不会提供类似的波幅,并且需要一组对角排列的正交电极对来产生足够波幅的信号,以便区分这些心律。

方法

在临床电生理实验室对40例植入植入式心脏复律除颤器(ICD)的患者进行研究。将一个由9个电极组成的方形阵列放置在患者左心上方的前胸,相邻电极间距为4至10厘米。最左边一列的中心电极位于左胸骨旁第5肋间。从8个外周电极相对于阵列中心电极进行记录,在SR以及诱发的VT/VF期间进行八通道记录。从这些记录中,通过代数方法构建所有20个相邻(包括对角相邻)电极的双极心电图。在每个双极导联以及通过对这些导联的每对相邻和正交导联的平方和求和形成的空间向量中,测量SR和VT/VF中的QRS峰峰值波幅。

结果

在SR中产生最佳心电图波幅的电极对,在VT/VF中并不总是产生最佳波幅的电极对。但在每位患者中都能找到两对正交电极,其QRS向量波幅在SR和VT/VF中都足以区分这些心律。

结论

一组由患者定义的双极电极对适用于在皮下动态监测中按心率自动分离SR和VT/VF。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验