Plagnard J, Morel J, Tuan A Tran
Laboratoire National Henri Becquerel, CEA/LNHB, 91191 Gif-sur-Yvette, Cedex, France.
Appl Radiat Isot. 2004 Feb-Apr;60(2-4):179-83. doi: 10.1016/j.apradiso.2003.11.013.
In gamma-ray spectrometry, new acquisition systems based on digital signal processing are now commercially available. In order to determine their performance at high count rates, the Commissariat à l'Energie Atomique (CEA), Laboratoire National Henri Becquerel has tested several of these systems. These tests have clearly shown that the performance levels announced by the manufacturers were generally not met. It was therefore a logical step to include the Atelier de DévelOppement Numérique pour l'Instrumentation en Spectrométrie (ADONIS) system in these tests. ADONIS is the new numerical system for gamma-ray spectrometry, developed by the CEA Service d'Instrumentation et d'Application des Rayonnements.
在伽马射线能谱分析中,基于数字信号处理的新型采集系统现已投入商业使用。为了确定这些系统在高计数率下的性能,法国原子能委员会(CEA)下属的亨利·贝克勒尔国家实验室对其中几种系统进行了测试。这些测试清楚地表明,制造商宣称的性能水平通常未能达到。因此,将数字仪器开发工作室(ADONIS)系统纳入这些测试是合乎逻辑的一步。ADONIS是由CEA辐射仪器与应用服务部开发的新型伽马射线能谱分析数字系统。