The measurement of the activity of therapeutic radiopharmaceuticals prior to the administration to patients is normally achieved via the use of radionuclide calibrators. An accurate measurement of the activity of pure beta-emitters is complex. Calibration problems can be solved by combining a primary calibration with a 90Y reference solution and a 90Sr/90Y transfer standard with a solid source, simulating geometric effects caused by high energetic beta radiation. The recent development of a 90Sr/90Y transfer standard for this purpose is reported.