Zeng G G, McEwen M R, Rogers D W, Klassen N V
National Research Council of Canada, Ottawa, K1A 0R6, Canada.
Phys Med Biol. 2004 Jan 21;49(2):257-70. doi: 10.1088/0031-9155/49/2/006.
The energy dependence of alanine/EPR dosimetry, in terms of absorbed dose-to-water for clinical 6, 10, 25 MV x-rays and 60Co gamma-rays was investigated by measurements and Monte Carlo (MC) calculations. The dose rates were traceable to the NRC primary standard for absorbed dose, a sealed water calorimetry. The electron paramagnetic resonance (EPR) spectra of irradiated pellets were measured using a Bruker EMX 081 EPR spectrometer. The DOSRZnrc Monte Carlo code of the EGSnrc system was used to simulate the experimental conditions with BEAM code calculated input spectra of x-rays and gamma-rays. Within the experimental uncertainty of 0.5%, the alanine EPR response to absorbed dose-to-water for x-rays was not dependent on beam quality from 6 MV to 25 MV, but on average, it was about 0.6% lower than its response to 60Co gamma-rays. Combining experimental data with Monte Carlo calculations, it is found that the alanine/EPR response per unit absorbed dose-to-alanine is the same for clinical x-rays and 60Co gamma-rays within the uncertainty of 0.6%. Monte Carlo simulations showed that neither the presence of PMMA holder nor varying the dosimeter thickness between 1 mm and 5 mm has significant effect of the energy dependence of alanine/EPR dosimetry within the calculation uncertainty of 0.3%.
通过测量和蒙特卡罗(MC)计算,研究了丙氨酸/电子顺磁共振剂量测定法对于临床6、10、25 MV X射线和60Co γ射线的水吸收剂量的能量依赖性。剂量率可追溯到美国国家研究委员会(NRC)吸收剂量的一级标准——密封水热法。使用布鲁克EMX 081电子顺磁共振光谱仪测量辐照颗粒的电子顺磁共振(EPR)光谱。EGSnrc系统的DOSRZnrc蒙特卡罗代码用于通过BEAM代码计算的X射线和γ射线输入光谱来模拟实验条件。在0.5%的实验不确定度范围内,丙氨酸EPR对X射线水吸收剂量的响应不依赖于6 MV至25 MV的射束质量,但平均而言,比其对60Co γ射线的响应低约0.6%。将实验数据与蒙特卡罗计算相结合,发现在0.6%的不确定度范围内,临床X射线和60Co γ射线每单位丙氨酸吸收剂量的丙氨酸/EPR响应是相同的。蒙特卡罗模拟表明,在0.3%的计算不确定度范围内,聚甲基丙烯酸甲酯(PMMA)支架的存在以及剂量计厚度在1 mm至
5 mm之间的变化对丙氨酸/电子顺磁共振剂量测定法的能量依赖性均无显著影响。