钍射气对脉冲电离室型氡探测器的影响。

Effects of thoron on a radon detector of pulse-ionization chamber type.

作者信息

Ishikawa T

机构信息

Radon Research Group, National Institute of Radiological Sciences, 4-9-1 Anagawa, Inage-ku, Chiba 263-8555, Japan.

出版信息

Radiat Prot Dosimetry. 2004;108(4):327-30. doi: 10.1093/rpd/nch037.

Abstract

A radon detector of pulse-ionization chamber (PIC) type could have some sensitivity for thoron. Thus, the presence of thoron could interfere with precise measurement of radon. In the present study, effects of thoron on the most common type of PIC detector (commercial name AlphaGUARD) were investigated using an exposure chamber. The AlphaGUARD was exposed to a mixture of radon and thoron, together with a radon/thoron discriminative monitor that employs a silicon solid-state detector. The thoron sensitivity of the PIC detector was estimated by comparing the two detectors. As a result, the thoron sensitivity was about 10% compared with the radon sensitivity. In other words, the radon concentration (Bq m(-3)) measured with the PIC detector was approximately the sum of the actual radon concentration (Bq m(-3)) and 10% of the thoron concentration (Bq m(-3)). The sensitivity to thoron should be considered in measurements in thoron-enhanced areas.

摘要

脉冲电离室(PIC)型氡探测器对钍射气可能有一定灵敏度。因此,钍射气的存在可能会干扰氡的精确测量。在本研究中,使用一个暴露室研究了钍射气对最常见类型的PIC探测器(商品名AlphaGUARD)的影响。将AlphaGUARD暴露于氡和钍射气的混合物中,同时还使用了一个采用硅固态探测器的氡/钍射气鉴别监测器。通过比较这两种探测器来估算PIC探测器对钍射气的灵敏度。结果表明,与对氡的灵敏度相比,对钍射气的灵敏度约为10%。换句话说,用PIC探测器测得的氡浓度(Bq m(-3))大约是实际氡浓度(Bq m(-3))与钍射气浓度(Bq m(-3))的10%之和。在钍射气增强区域进行测量时,应考虑对钍射气的灵敏度。