Lee K B, Lee Jong Man, Park Tae Soon
Korea Research Institute of Standards and Science, P.O. Box 102, Yuseong, Daejeon 305-600, South Korea.
Appl Radiat Isot. 2004 Jun;60(6):893-7. doi: 10.1016/j.apradiso.2004.01.013.
The CIEMAT/NIST efficiency tracing method has been widely used for radionuclide standardization. The method consists of the theoretical calculations of counting efficiencies and the experimental data obtained from a commercially available double photomultiplier detector. We have recently implemented the technique using (3)H as a tracer. Our implementation of the method was validated by participating into an international comparison of activity measurements of a solution of (204)T1. We report in this paper the procedure employed to implement the technique and compare our measurement result of (204)T1 activity with that of the international comparison.