Elsila Jamie E, de Leon Nathalie P, Zare Richard N
Department of Chemistry, Stanford University, Stanford, California 94305-5080, USA.
Anal Chem. 2004 May 1;76(9):2430-7. doi: 10.1021/ac0354140.
Microprobe laser desorption/laser ionization mass spectrometry (microL(2)MS) is a sensitive and selective technique that has proven useful in the qualitative and semiquantitative detection of trace organic compounds, particularly polycyclic aromatic hydrocarbons (PAHs). Recent efforts have focused on developing microL(2)MS as a quantitative method, often by measuring the ratio of signal strength of an analyte to an internal standard. Here, we present evidence of factors that affect these ratios and thus create uncertainty and irreproducibility in quantification. The power and wavelength of the desorption laser, the delay time between the desorption and ionization steps, the power of the ionization laser, and the ionization laser alignment are all shown to change PAH ratios, in some cases by up to a factor of 24. Although changes in the desorption laser parameters and the delay time cause the largest effects, the ionization laser power and alignment are the most difficult parameters to control and thus provide the most practical limitations for quantitative microL(2)MS. Variation in ratios is seen in both synthetic poly(vinyl chloride) membranes and in "real-life" samples of Murchison meteorite powder. Ratios between similar PAHs vary less than those between PAHs that differ greatly in mass and structure. This finding indicates that multiple internal standards may be needed for quantification of samples containing diverse PAHs.
微探针激光解吸/激光电离质谱法(microL(2)MS)是一种灵敏且具有选择性的技术,已被证明在痕量有机化合物,特别是多环芳烃(PAHs)的定性和半定量检测中很有用。最近的工作重点是将microL(2)MS发展成为一种定量方法,通常是通过测量分析物与内标物的信号强度比来实现。在此,我们展示了影响这些比例的因素的证据,从而在定量分析中产生不确定性和不可重复性。解吸激光的功率和波长、解吸与电离步骤之间的延迟时间、电离激光的功率以及电离激光的对准情况均显示会改变PAH比例,在某些情况下变化幅度高达24倍。尽管解吸激光参数和延迟时间的变化产生的影响最大,但电离激光的功率和对准情况是最难控制的参数,因此为定量microL(2)MS提供了最实际的限制。在合成聚氯乙烯膜和默奇森陨石粉末的“实际”样品中都观察到了比例的变化。质量和结构差异较大的PAHs之间的比例变化比相似PAHs之间的比例变化更大。这一发现表明,对于含有多种PAHs的样品进行定量分析可能需要多个内标物。