Koal Therese, Deters Michael, Casetta Bruno, Kaever Volkhard
Medical School Hannover, Institute of Pharmacology, D-30625, Germany.
J Chromatogr B Analyt Technol Biomed Life Sci. 2004 Jun 15;805(2):215-22. doi: 10.1016/j.jchromb.2004.02.040.
In this study immunosuppressants, i.e. cyclosporin A (CyA), tacrolimus (TRL), sirolimus (SRL) and everolimus (RAD) were quantified in whole blood samples from immunosuppressant treated transplant recipients by an integrated on-line solid phase extraction-high performance liquid chromatography-tandem mass spectrometry (SPE-HPLC-MS/MS) system. This method has been developed to improve the following characteristics: speed, robust analysis, simultaneous determination and low cost. This can be achieved by the use of a perfusion column as an extraction cartridge in combination with a short HPLC column and highly selective and sensitive atmospheric pressure ionisation tandem mass spectrometry (API-MS/MS) in the multiple reaction monitoring (MRM) detection mode. This high throughput technique is perfectly appropriate for routine therapeutic drug monitoring (TDM) of organ transplanted patients.