Zhang Feng, Chen Zhencheng, Deng Zhensheng, Jiang Dazong
Institute of Biomedical Engineering, Xi'an Jiaotong University, Xi'an 710049, China.
Sheng Wu Yi Xue Gong Cheng Xue Za Zhi. 2004 Apr;21(2):280-3.
Dual-kilovolt technique was realized in dual energy X-ray imaging system with computer controlled KV adjustment, filter switching and image acquirement. The dual energy images can be obtained in 2-3 seconds to avoid moving artifact. Low energy photons in high energy X-ray beam can be attenuated effectively by a 2 mm Cu filter to separate the spectra of the high and low energy X-rays. A simplified dual energy decomposition method was used to obtain the soft tissue image. An example of chest dual energy radiographs is presented to illustrate the performance of the system.
通过计算机控制千伏调节、滤过器切换和图像采集,在双能X射线成像系统中实现了双千伏技术。可在2至3秒内获取双能图像,以避免运动伪影。高能X射线束中的低能光子可被2毫米铜滤过器有效衰减,从而分离高低能X射线的光谱。采用一种简化的双能分解方法来获取软组织图像。给出了胸部双能X线片的示例以说明该系统的性能。