A systematic evaluation of bioelectrical impedance measurement after hemodialysis session.

作者信息

Di Iorio Biagio R, Scalfi Luca, Terracciano Vincenzo, Bellizzi Vincenzo

机构信息

Nephrology and Dialysis Unit, Solofra Hospital, Italy.

出版信息

Kidney Int. 2004 Jun;65(6):2435-40. doi: 10.1111/j.1523-1755.2004.00660.x.

Abstract

BACKGROUND

There is still no definitive indication about the ideal point of time to perform bioimpedance analysis (BIA) in hemodialysis patients. Furthermore, the interpretation of data in this regard is difficult because there is still no comprehensive information about the fluctuations in BIA variables occurring in these subjects. The aim of this study was to assess BIA changes occurring in hemodialysis and specifically in the dry-weight state.

METHODS

We studied 27 anuric patients (20 males and 7 females; age 56.1 +/- 13.7 years) on chronic hemodialysis. Single-frequency BIA (R, resistance; Xc, reactance; and PhA, phase angle) was performed (1) before and at the end of hemodialysis (dialysis period); (2) 15, 30, 60, 90, and 120 minutes after hemodialysis (postdialysis period); and (3) 24, 48, and 68 hours after hemodialysis (interdialysis period).

RESULTS

Body weight decreased by 2.8 +/- 0.8 kg during hemodialysis, was unchanged during the postdialysis period, and progressively rose during the interdialysis period. At the same time, BIA variables significantly increased during hemodialysis (R, 453 +/- 74 and 542 +/- 98 ohm; Xc, 38 +/- 10 and 53 +/- 16 ohm; P < 0.05), remained stable over the 120-minute period after treatment (R, 538 +/- 94, 539 +/- 95, 538 +/- 94, 541 +/- 95, and 544 +/- 95 ohm; and Xc, 53 +/- 15, 53 +/- 15, 51 +/- 16, 52 +/- 16, and 52 +/- 16 ohm; NS), and subsequently declined [R, 471 +/- 79 (P= <0.05 vs. postdialysis), 449 +/- 71 (P= <0.05 vs. postdialysis), 424 +/- 68 (P= <0.05 vs. postdialysis) ohm; Xc, 42 +/- 13 (P= <0.05 vs. postdialysis), 37 +/- 10 (P= <0.05 vs. postdialysis), 34 +/- 13 (P= <0.05 vs. postdialysis) ohm]. The stability of BIA measures during postdialysis was confirmed by the constant relationship found between R/height and Xc/height. Also PhA increased after dialysis (4.8 +/- 1.1 degrees vs. 5.7 +/- 1.3 degrees, P < 0.05), was unchanged during the following 120 minutes and decreased in the interdialysis period (5.1 +/- 1.3 degrees, 4.8 +/- 1.0 degrees, and 4.5 +/- 1.1 degrees, P < 0.05). At the end of hemodialysis and during the postdialysis period total body water (TBW) estimated from BIA was similar on average to TBW calculated using Watson formulas (37.2 +/- 6.3 L vs. 36.2 +/- 5.7 L, NS). On the contrary, when patients were hyperhydrated BIA significantly overestimated the Watson's values.

CONCLUSION

In hemodialysis patients BIA variables fluctuate to a considerable extent (with the highest values immediately after hemodialysis), but remain constant and highly reproducible over the 120 minutes after the end of hemodialysis, that is, in a dry-weight state. Thus, taking into consideration that the point in time chosen for performing BIA is crucial to properly assess body composition, BIA can be appropriately performed at anytime during the postdialysis period, provided that hydration status does not change due to food or drink consumption.

摘要

背景

对于血液透析患者进行生物电阻抗分析(BIA)的理想时间点仍没有明确的指征。此外,由于仍没有关于这些受试者BIA变量波动的全面信息,因此在这方面对数据的解读很困难。本研究的目的是评估血液透析患者,特别是干体重状态下BIA的变化。

方法

我们研究了27例慢性血液透析的无尿患者(20例男性和7例女性;年龄56.1±13.7岁)。进行单频BIA(R,电阻;Xc,电抗;和PhA,相位角)测量:(1)在血液透析前和结束时(透析期);(2)血液透析后15、30、60、90和120分钟(透析后期);(3)血液透析后24、48和68小时(透析间期)。

结果

血液透析期间体重下降2.8±0.8kg,透析后期体重无变化,透析间期体重逐渐上升。同时,BIA变量在血液透析期间显著增加(R,453±74和542±98欧姆;Xc,38±10和53±16欧姆;P<0.05),在治疗后的120分钟内保持稳定(R,538±94、539±95、538±94、541±95和544±95欧姆;Xc,53±15、53±15、51±16、52±16和52±16欧姆;无显著性差异),随后下降[R,471±79(与透析后相比P<0.05)、449±71(与透析后相比P<0.05)、424±68(与透析后相比P<0.05)欧姆;Xc,42±13(与透析后相比P<0.05)、37±10(与透析后相比P<0.05)、34±13(与透析后相比P<0.05)欧姆]。透析后期BIA测量的稳定性通过R/身高和Xc/身高之间的恒定关系得到证实。此外,PhA在透析后增加(4.8±1.1度对5.7±1.3度,P<0.05),在随后的120分钟内无变化,在透析间期下降(5.1±1.3度、4.8±1.0度和4.5±1.1度,P<0.05)。在血液透析结束时和透析后期,通过BIA估计的全身水(TBW)平均与使用Watson公式计算的TBW相似(37.2±6.3L对36.2±5.7L,无显著性差异)。相反,当患者水合过多时,BIA显著高估了Watson的值。

结论

在血液透析患者中,BIA变量波动较大(血液透析后立即达到最高值),但在血液透析结束后的120分钟内,即在干体重状态下保持恒定且具有高度可重复性。因此,考虑到进行BIA所选的时间点对于正确评估身体成分至关重要,只要水合状态不因食物或饮料摄入而改变,在透析后期的任何时间都可以适当地进行BIA。