对一个在空气中用1000千戈瑞伽马射线交联的超高分子量聚乙烯髋臼杯的分析。

Analysis of a retrieved UHMWPE acetabular cup crosslinked in air with 1000 kGy of gamma radiation.

作者信息

Sugano N, Saito M, Yamamoto T, Nishii T, Yau S-S, Wang A

机构信息

Department of Orthopaedic Surgery, Osaka University Graduate School of Medicine, 2-2 Yamadaoka, Suita 565-0871, Japan.

出版信息

J Orthop Res. 2004 Jul;22(4):828-31. doi: 10.1016/j.orthres.2003.12.001.

Abstract

The oxidation and degradation that accompany high dose gamma irradiation in air for crosslinking and sterilization are important because they could affect the clinical performance of polyethylene total joint implants. We report a clinical case of a 1000 kGy gamma-irradiated, highly crosslinked ultra-high molecular weight polyethylene acetabular cup retrieved 24 years postoperatively. Performance evaluations included absorbed radiation dose, total wear penetration, average wear rate, crystallinity, oxidation, and equibiaxial tensile properties. A retrieved acetabular cup of the same grade of polyethylene but gamma-sterilized using a conventional dose was used as control. The highly crosslinked and control cups took in about 1300 and 30 kGy of radiation, respectively, as measured using a trans-vinylene index. Direct dimensional measurements revealed average wear rates of the highly crosslinked and control cups were 0.04 and 0.06 mm/year, respectively. The oxidation index of the highly crosslinked cup was very high (0.679), but equivalent to that of a 1000 kGy irradiated reference cup. The retrieved highly crosslinked cup showed much higher equibiaxial ultimate tensile strength than the retrieved control cup. Based on these observations, the increased wear resistance and equibiaxial tensile properties that resulted from extensive crosslinking in the presence of air were partially offset by the adverse effects caused by immediate oxidation during the process.

摘要

在空气中进行交联和灭菌的高剂量伽马辐射所伴随的氧化和降解过程很重要,因为它们可能会影响聚乙烯全关节植入物的临床性能。我们报告了一例临床病例,该病例为一个在术后24年取出的、经1000千戈瑞伽马辐射的、高度交联的超高分子量聚乙烯髋臼杯。性能评估包括吸收辐射剂量、总磨损穿透量、平均磨损率、结晶度、氧化程度以及双轴拉伸性能。使用一个相同等级但采用传统剂量进行伽马灭菌的聚乙烯髋臼杯作为对照。通过反式亚乙烯基指数测量发现,高度交联的髋臼杯和对照杯分别吸收了约1300千戈瑞和30千戈瑞的辐射。直接尺寸测量显示,高度交联的髋臼杯和对照杯的平均磨损率分别为每年0.04毫米和0.06毫米。高度交联髋臼杯的氧化指数非常高(0.679),但与一个经1000千戈瑞辐射的参考杯相当。取出的高度交联髋臼杯显示出比取出的对照杯高得多的双轴极限拉伸强度。基于这些观察结果,在空气中进行广泛交联所带来的耐磨性和双轴拉伸性能的提高,在一定程度上被该过程中即时氧化所造成的不利影响抵消。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验