Suppr超能文献

使用微通道板探测器的扫描狭缝光子计数X射线成像系统。

Scanning-slit photon counting x-ray imaging system using a microchannel plate detector.

作者信息

Shikhaliev Polad M, Xu Tong, Le Huy, Molloi Sabee

机构信息

Imaging Physics Laboratory, Department of Radiological Sciences, University of California, Irvine, California 92697, USA.

出版信息

Med Phys. 2004 May;31(5):1061-71. doi: 10.1118/1.1695651.

Abstract

An experimental prototype of a novel photon counting x-ray imaging system was evaluated. This system is based on an "edge-on" microchannel plate (MCP) detector and utilizes scanning slit imaging configuration. The detector is capable of photon counting, direct conversion, high spatial resolution, controllable physical charge amplification, quantum limited and scatter free operation. The detector provides a 60 mm wide field of view (FOV) and its count rate is 200 kHz for the entire FOV. The count rate of the current system is limited by the position encoding electronics, which has a single input for all events from the entire detector, and incorporates a single channel ADC with 1 micros conversion time. It is shown that the count rate can potentially be improved to clinically acceptable levels using multichannel application specific integrated circuit (ASIC) electronics and multi-slit image acquisition geometry. For a typical acquisition time used in this study, the image noise was measured to be less than the typically acceptable noise level for medical x-ray imaging. It is anticipated that the noise level will be also low after the implementation of the ASIC electronics. The quantum efficiency of the detector was measured to be 40%-56% for an energy range of 50-90 kVp for MCPs used in this study and can be improved to > 80% using MCPs with the optimized parameters. Images of resolution and anthropomorphic phantoms were acquired at an x-ray tube voltage of 50 kVp. The value of contrast transfer function for the detector was measured to be 0.5 at a spatial frequency of 5 lp/mm. The intrinsic spatial resolution of the system is 28 microm FWHM and was limited by the accuracy of the time-to-digital conversion of the position encoding electronics. Given the advantages of the edge-on MCP detector such as direct conversion and physical charge amplification, it can potentially be applied to mammography and chest radiography.

摘要

对一种新型光子计数X射线成像系统的实验原型进行了评估。该系统基于“边缘入射”微通道板(MCP)探测器,并采用扫描狭缝成像配置。该探测器能够进行光子计数、直接转换、高空间分辨率、可控的物理电荷放大、量子极限且无散射操作。探测器提供60毫米宽的视场(FOV),其整个视场的计数率为200千赫兹。当前系统的计数率受位置编码电子设备的限制,该设备对来自整个探测器的所有事件只有一个输入,并包含一个转换时间为1微秒的单通道模数转换器(ADC)。结果表明,使用多通道专用集成电路(ASIC)电子设备和多狭缝图像采集几何结构,计数率有可能提高到临床可接受的水平。对于本研究中使用的典型采集时间,测量得到的图像噪声低于医学X射线成像通常可接受的噪声水平。预计在实施ASIC电子设备后,噪声水平也会很低。对于本研究中使用的MCP,在50 - 90 kVp的能量范围内,探测器的量子效率测量值为40% - 56%,使用具有优化参数的MCP可将其提高到> 80%。在X射线管电压为50 kVp时采集了分辨率和模拟人体模型的图像。在空间频率为5线对/毫米时,探测器的对比度传递函数值测量为0.5。系统的固有空间分辨率为28微米半高宽(FWHM),并受位置编码电子设备时间数字转换精度的限制。鉴于边缘入射MCP探测器的优点,如直接转换和物理电荷放大,它有可能应用于乳腺摄影和胸部X线摄影。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验