Measurement of forces during laryngoscopy.

作者信息

Bucx M J, Scheck P A, Van Geel R T, Den Ouden A H, Niesing R

机构信息

Department of Anaesthesia, Erasmus University Rotterdam, The Netherlands.

出版信息

Anaesthesia. 1992 Apr;47(4):348-51. doi: 10.1111/j.1365-2044.1992.tb02180.x.

Abstract

The forces transmitted by the laryngoscope blade onto the base of the tongue are assumed to be a major cardiovascular stimulus. This study investigates the various forces applied onto a Macintosh laryngoscope during laryngoscopy. The forces acting along the axis of the handle are described, as well as the forces exerted by the anaesthetist to prevent the laryngoscope from turning. Based on a relatively simple technique, a modified laryngoscope has been designed to measure these forces. Four different laryngoscopic parameters are determined: (1) the duration of laryngoscopy, (2) the maximally applied force, (3) the mean force and (4) the integral of the force over the time (area under the curve). The force measurements of 49 anatomically normal patients undergoing uncomplicated intubation are included in the study. The duration of laryngoscopy was 16.3s (SD 11.8), the applied peak force was 35N (SD 12) and mean force necessary was 20N (SD 6) while the force-time integral was 324N (SD 194).

摘要

喉镜叶片作用于舌根的力量被认为是主要的心血管刺激因素。本研究调查了喉镜检查过程中作用于麦金托什喉镜上的各种力量。描述了沿手柄轴线方向的作用力,以及麻醉师为防止喉镜转动而施加的力。基于一种相对简单的技术,设计了一种改良喉镜来测量这些力量。确定了四个不同的喉镜参数:(1)喉镜检查持续时间,(2)最大施加力,(3)平均力,(4)力随时间的积分(曲线下面积)。该研究纳入了49例接受简单插管且解剖结构正常患者的力量测量数据。喉镜检查持续时间为16.3秒(标准差11.8),施加的峰值力为35牛(标准差12),所需平均力为20牛(标准差6),而力-时间积分为324牛(标准差194)。