Ochiai R, Nagano M
Department of Anesthesiology, School of Medicine Keio University, Tokyo, Japan.
J Anesth. 1990 Jan;4(1):61-5. doi: 10.1007/s0054000040061.
The effect of calcium-balanced heparin (471896, CIBA CORNING) on blood gas and electrolyte analysis was evaluated, by comparing with that of sodium heparin (Na heparin). One ml of whole blood was collected into a syringe, which contained calcium-balance heparin (Ca balanced heparin) or Na heparin. 122 pairs of blood samples obtained from 15 patients were analyzed for Na, K, ionized calcium (Ca(++)), total hemoglobin, pH, P(CO)(2), and P(O)(2) by an automatic blood gas and electrolyte analyzer, CIBA CORNING model 288. There was a significant difference ( P << 0.05) in pH, P(CO)(2), Na, and Ca(++) between the two different groups. Ca(++) concentration was significantly less in Na heparin group than in Ca balanced heparin group, probably due to more chelation of Ca(++) by Na heparin than Ca balanced heparin. The present study suggests that the Ca balanced heparin has minimal effect on the blood gas and electrolyte analysis, and is a suitable anticoagulant for the Ca(++) measurement.