Yu Hua, Yan Feng, Dai Zong, Ju Huangxian
State Key Laboratory of Coordination Chemistry, Department of Chemistry, Institute of Chemical Biology, Nanjing University, Nanjing 210093, People's Republic of China.
Anal Biochem. 2004 Aug 1;331(1):98-105. doi: 10.1016/j.ab.2004.03.042.
A screen-printed three-electrode system is fabricated to prepare a novel disposable screen-printed immunosensor for rapid determination of alpha-1-fetoprotein (AFP) in human serum. The immunosensor is prepared by entrapping horseradish peroxidase (HRP)-labeled AFP antibody in chitosan membrane to modify the screen-printed carbon electrode. The membrane is characterized with scanning electron microscope and electrochemical methods. After the immunosensor is incubated with AFP at 30 degrees C for 35 min, the access of the active center of HRP catalyzing the oxidation reaction of thionine by H(2)O(2) is partly inhibited. In presence of 1.2 mM thionine and 6 mM H(2)O(2), the electrocatalytic current decreases linearly in two concentration ranges of AFP from 0 to 20 and from 20 to 150 ng/mL with a detection limit of 0.74 ng/mL. The immunosensor shows an acceptable accuracy compared with those obtained from immunoradiometric assays. The interassay coefficients of variation are 6.6 and 4.2% at 10 and 100 ng/mL, respectively. The storage stability is acceptable in pH 7.0 phosphate buffer solution at 4 degrees C for more than 10 days. The proposed method can detect the AFP through one-step immunoassay and would be valuable for clinical immunoassay.
制备了一种丝网印刷三电极系统,以制备一种新型一次性丝网印刷免疫传感器,用于快速测定人血清中的甲胎蛋白(AFP)。该免疫传感器是通过将辣根过氧化物酶(HRP)标记的AFP抗体包埋在壳聚糖膜中来修饰丝网印刷碳电极而制备的。用扫描电子显微镜和电化学方法对该膜进行了表征。免疫传感器在30℃下与AFP孵育35分钟后,HRP催化硫堇被H₂O₂氧化反应的活性中心的可及性受到部分抑制。在1.2 mM硫堇和6 mM H₂O₂存在下,AFP在0至20和20至150 ng/mL两个浓度范围内,电催化电流呈线性下降,检测限为0.74 ng/mL。与免疫放射分析相比,该免疫传感器具有可接受的准确性。在10和100 ng/mL时,批间变异系数分别为6.6%和4.2%。在4℃的pH 7.0磷酸盐缓冲溶液中,储存稳定性良好,可保存10天以上。所提出的方法可以通过一步免疫测定法检测AFP,对临床免疫测定具有重要价值。