Ablation lesion size correlates with pacing threshold: a physiological basis for use of pacing to assess ablation lesions.

作者信息

Sapp John L, Soejima Kyoko, Cooper Joshua M, Epstein Laurence M, Stevenson William G

机构信息

Division of Cardiology, Department of Medicine, QEII Health Sciences Centre, Dalhousie University, Halifax, Canada.

出版信息

Pacing Clin Electrophysiol. 2004 Jul;27(7):933-7. doi: 10.1111/j.1540-8159.2004.00561.x.

Abstract

The virtual electrode model predicts that pacing stimulus strength should reflect proximity of the pacing electrode to excitable myocardium, allowing pacing threshold to assess radiofrequency (RF) ablation lesions and unexcitable scar. The purpose of this study is to correlate RF lesion size with pacing threshold and electrogram (EG) amplitude change at the ablation site. In four swine (32-58 kg, 20 ventricular RF lesions were created using a 4-mm tip electrode catheters under fluoroscopic and electroanatomic guidance. Unipolar pacing threshold and bipolar and unipolar EG amplitude were measured before and after ablation and compared with lesion size measured in the fixed, serially sectioned tissue. Lesion diameter ranged from 6.4 to 19 mm and volume ranged from 29 to 1920 mm3. Ablation increased the pacing threshold by 320%, from 0.9 +/- 0.3 to 3.6 +/- 2.6 mA, P < 0.001. The change in pacing threshold correlated with lesion volume R = 0.88, P < 0.001). Linear regression predicts that lesion volume (mm3) = 160 X rise in pacing threshold + 13. Ablation reduced peak to peak bipolar EG amplitude by 56%, from 2.5 +/- 2.0 mV to 1.1 +/- 0.6 mV (P = 0.005). Unipolar EG amplitude diminished by only 22% from 4.0 +/- 1.6 to 3.2 +/- 0.9 mV postablation (P = 0.005). The correlations of lesion volume with change in either bipolar R = 0.14, P = 0.6) or unipolar R = 0.18, P = 0.6) EG amplitude were poor. Pacing threshold correlates with RF ablation lesion size, consistent with the virtual electrode model. In normal myocardium, change in pacing threshold is likely to be a better marker of lesion size than electrogram amplitude.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验