Banks Scott A, Hodge W Andrew
The Biomotion Foundation, West Palm Beach, FL 33401, USA.
Clin Orthop Relat Res. 2004 Sep(426):187-93. doi: 10.1097/01.blo.0000138956.04316.ac.
Knee implant motions have a direct influence on patient function and implant longevity. The purpose of this study was to determine if there were consistent differences in knee motions among three groups of knee implants. Two hundred thirteen knees in 173 patients, with 25 implant designs, were studied using fluoroscopy during stair-stepping. All knee implants were assigned to one of three groups based on the design: fixed-bearing posterior-stabilized, fixed-bearing posterior cruciate-retaining, and mobile-bearing. All types of implants had the same pattern of internal/external rotations, but different designs had different anteroposterior translations. Seventy-five percent of posterior-stabilized knee implants had a medial center of rotation, indicating posterior femoral translation with flexion. Sixty-three percent of cruciate-retaining fixed-bearing knee implants had a lateral center of rotation. Eighty-six percent of mobile-bearing knee implants had a lateral center of rotation, indicating anterior femoral translation with flexion. Knee motion in patients with successful total knee arthroplasties is related directly to the constraints of the implant design.
膝关节植入物的运动对患者功能和植入物使用寿命有直接影响。本研究的目的是确定三组膝关节植入物的膝关节运动是否存在一致差异。对173例患者的213个膝关节、25种植入物设计进行了研究,在上下楼梯过程中使用荧光透视法。根据设计,所有膝关节植入物被分为三组之一:固定平台后稳定型、固定平台后交叉韧带保留型和活动平台型。所有类型的植入物都有相同的内/外旋转模式,但不同设计的前后平移不同。75%的后稳定型膝关节植入物有内侧旋转中心,表明股骨后移伴屈曲。63%的后交叉韧带保留型固定平台膝关节植入物有外侧旋转中心。86%的活动平台型膝关节植入物有外侧旋转中心,表明股骨前移伴屈曲。全膝关节置换术成功患者的膝关节运动与植入物设计的限制直接相关。