Tanimura Y, Yoshizawa M, Saegusa J, Fujii K, Shimizu S, Yoshida M, Shibata Y, Uritani A, Kudo K
Japan Atomic Energy Research Institute, Tokai-mura, Naka-gun, Ibaraki, 319-1195, Japan.
Radiat Prot Dosimetry. 2004;110(1-4):85-9. doi: 10.1093/rpd/nch197.
Monoenergetic neutron calibration fields of 144, 565 keV and 5.0 MeV have been developed at the Facility of Radiation Standards of JAERI using a 4 MV Pelletron accelerator. The 7Li(p,n)7Be and 2H(d,n)3He reactions are employed for neutron production. The neutron energy was measured by the time-of-flight method with a liquid scintillation detector and calculated with the MCNP-ANT code. A long counter is employed as a neutron monitor because of the flat response. The monitor is set up where the influence of inscattered neutrons from devices and their supporting materials at a calibration point is as small as possible. The calibration coefficients from the monitor counts to the neutron fluence at a calibration point were obtained from the reference fluence measured with the transfer instrument of the primary standard laboratory (AIST), a 24.13 cm phi Bonner sphere counter. The traceability of the fields to AIST was established through the calibration.
日本原子能研究所辐射标准设施利用一台4兆伏的佩莱顿加速器,开发出了能量分别为144千电子伏、565千电子伏和5.0兆电子伏的单能中子校准场。利用7Li(p,n)7Be和2H(d,n)3He反应来产生中子。中子能量通过飞行时间法,使用液体闪烁探测器进行测量,并通过MCNP-ANT代码进行计算。由于响应平坦,采用长计数器作为中子监测器。将监测器设置在校准点处,使来自设备及其支撑材料的散射中子在校准点的影响尽可能小。校准点处从监测器计数到中子注量的校准系数,是根据用一级标准实验室(日本国家计量院)的传递仪器——一个直径24.13厘米的邦纳球计数器测量的参考注量获得的。通过校准建立了这些场与日本国家计量院的溯源性。