Summers Paul E, Kollias Spyros S, Valavanis Anton
Institute of Neuroradiology, University Hospital Zurich, Zurich, Switzerland.
J Magn Reson Imaging. 2004 Oct;20(4):662-73. doi: 10.1002/jmri.20156.
To evaluate the use of parallel imaging (sensitivity encoding [SENSE]) to improve spatial resolution and achieve sub-second temporal resolution in fluoroscopic contrast-enhanced, magnetic resonance digital subtraction angiography (MR-DSA).
A MR-DSA sequence was optimized on a 3-T scanner with respect to sampling bandwidth and SENSE acceleration factor subject to the constraints of half-second acquisition time and 0.6 x 1.2 mm in-plane resolution. MR-DSA with and without SENSE acceleration was then evaluated in patients with arterio-venous malformations (AVMs).
Consistent with previously reported results and theory, SENSE factors greater than two and increasing sampling bandwidth both led to increasing image noise. Compared to lower resolution MR-DSA images with similar temporal resolution, the SENSE accelerated sequence provided better spatial resolution without notable changes in the contrast enhancement of the vascular territories of the AVMs but was hampered somewhat in the late venous phases by a reconstruction artifact.
SENSE acceleration of MR-DSA by a factor of two allows improved temporal or spatial resolution without significant loss of image quality. Signal-to-noise degradation associated with higher SENSE acceleration factors are likely to necessitate other approaches to further improving resolution in MR-DSA. Clinically, SENSE accelerated MR-DSA improves the non-invasive pre- and postoperative depiction of AVM flow dynamics.
评估在荧光透视对比增强磁共振数字减影血管造影(MR-DSA)中使用并行成像(敏感性编码[SENSE])来提高空间分辨率并实现亚秒级时间分辨率。
在一台3-T扫描仪上,在采集时间为半秒且平面分辨率为0.6×1.2 mm的限制条件下,针对采样带宽和SENSE加速因子对MR-DSA序列进行优化。然后在患有动静脉畸形(AVM)的患者中评估有无SENSE加速的MR-DSA。
与先前报道的结果和理论一致,大于2的SENSE因子以及增加的采样带宽均导致图像噪声增加。与具有相似时间分辨率的较低分辨率MR-DSA图像相比,SENSE加速序列提供了更好的空间分辨率,AVM血管区域的对比增强没有明显变化,但在静脉晚期受到重建伪影的一定影响。
将MR-DSA的SENSE加速因子提高两倍可改善时间或空间分辨率,而不会显著损失图像质量。与更高的SENSE加速因子相关的信噪比下降可能需要其他方法来进一步提高MR-DSA的分辨率。临床上,SENSE加速的MR-DSA可改善AVM血流动力学的无创术前和术后描绘。