Prostate dosimetry in an anthropomorphic phantom.

作者信息

Yu P K N, Cheung T, Butson M J

机构信息

City University of Hong Kong, Department of Physics and Materials Science, Kowloon Tong, Hong Kong.

出版信息

Australas Phys Eng Sci Med. 2004 Jun;27(2):60-2. doi: 10.1007/BF03178377.

Abstract

Four field prostate treatments are a standard treatment procedure in radiotherapy. Dose in the prostate and rectum region were calculated for 6MV and 18MV photon beams on an anthropomorphic phantom with a collapsed cone convolution method using a 3-D planning system. Validation has been performed with radiographic film and thermoluminescent dosimeters. Results have shown that the pinnacle planning system has accurately modelled doses delivered to a heterogeneous phantom with calculations and measurements agreeing within +/-3% over most areas. When treating clinically, considerations such as the volume of bowel gas should be taken into account when planning. A sample of patient CT scans showed that in the absence of a heterogeneity correction, the error in estimated dose through the rectum could be as high as 8% in the presence of large volumes of rectal gas. Considerations, such as whether the patient undergoes another CT scan, the bowel gas volume ignored or assigned a specific density needs to be taken into account and brought to the attention of the radiation oncologists for accurate treatment.