Extensive coagulation monitoring in patients after implantation of the MicroMed Debakey continuous flow axial pump.

作者信息

Bonaros Nikolaos, Mueller Michael-Rolf, Salat Andreas, Schima Heinrich, Roethy Wilfried, Kocher Alfred A, Wolner Ernst, Wieselthaler Georg M

机构信息

Department of Cardiothoracic Surgery, University of Vienna, Austria.

出版信息

ASAIO J. 2004 Sep-Oct;50(5):424-31. doi: 10.1097/01.mat.0000136515.97686.a2.

Abstract

Ventricular assist device (VAD) implantation is associated with impaired primary hemostasis and thromboembolic complications. Recently, a new generation of implantable continuous flow axial pumps was introduced into clinical application. To study the potential thrombogenic properties of this type of pump, we applied extensive platelet monitoring was applied. In our institution, 13 patients received the MicroMed DeBakey VAD as a bridge to transplantation. Routine coagulation tests (platelet count, activated partial thromboplastin time, prothrombin time, antithrombin III activity) and platelet function tests (whole blood aggregometry, thrombelastography, flow cytometry) were performed. No clinically relevant thromboembolic events were detected. No correlation was found between global function tests, platelet aggregation, and thrombelastography. No correlation was detected between platelet activation and hemolysis parameters. Platelet aggregation and coagulation index were significantly suppressed early after operation. A subsequent phase of hyper-aggregability, starting around day 6, suggested the initiation of antiaggregation therapy. Platelet activation markers were upregulated in the postoperative period but were returned to preoperative levels after initiation of aspirin. In contrast to routine coagulation monitoring, platelet function tests reflect in detail the coagulation status of blood pump recipients and the efficiency of antiaggregation therapy. Aspirin and dipyridamole therapy in addition to oral anticoagulation using phenprocoumon may contribute to platelet function and clot mechanics restoration and is, therefore, recommended for patients after VAD implantation.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验