Increased digitalis-like immunoreactive substances in neonatal plasma measured using fluorescence polarization immunoassay.

作者信息

Ijiri Y, Hayahi T, Ogihara T, Ohi K, Suzuki K, Tamai H, Kitaura Y, Takenaka H, Tanaka K

机构信息

Department of Pharmacy, Osaka Medical College Hospital, Osaka, Japan.

出版信息

J Clin Pharm Ther. 2004 Dec;29(6):565-71. doi: 10.1111/j.1365-2710.2004.00609.x.

Abstract

OBJECTIVE

To better define the reported increased digitalis-like immunoreactive substances (DLIS) in neonatal plasma, we studied the relation among plasma DLIS level, blank intensity (BLK-I) value at FPIA measurement and plasma total bilirubin level.

METHODS

The DLIS levels were measured in 10 neonates with or without jaundice and 10 infants in good health, using fluorescence polarization immunoassay (FPIA) and microparticle enzyme immunoassay (MEIA). BLK-I value and plasma total bilirubin level were also measured simultaneously.

RESULTS

In neonates with jaundice, DLIS using FPIA, BLK-I and total bilirubin level were 0.58 +/-0.13 ng/mL, 2598 +/- 408, and 17.98 +/- 1.13 mg/dL, respectively, before phototherapy, and 0.33 +/-0.06 ng/mL, 1886 +/- 237, and 15.16 +/- 2.07 mg/dL after phototherapy. Corresponding values in neonates without jaundice were (DLIS: 0.34 +/-0.04 ng/mL; BLK-I: 1,764 +/- 278; total bilirubin: 10.37 +/- 4.54 mg/dL); in healthy infants (0.12 +/-0.06 ng/mL, 400.7 +/- 4.6 and 0.42 +/- 0.13 mg/dL, respectively) and in healthy volunteers (0.10 +/-0.07 ng/mL, 403.1 +/- 8.4, and 0.58 +/- 0.30 mg/dL, respectively). Using MEIA, DLIS was not detected in 10 neonates, 10 infants and 20 healthy volunteers.

CONCLUSIONS

A fluorescent compound related to bilirubin increased the BLK-I value in the measurement of neonatal plasma using FPIA. The fluorescence was not the result of endogenous digitalis-like factors.