Coedo A G, Padilla I, Dorado M T
Centro Nacional de Investigaciones Metalúrgicas (CSIC), Avda. Gregorio del Amo, 8 28040 Madrid, Spain.
Appl Spectrosc. 2004 Dec;58(12):1481-7. doi: 10.1366/0003702042641371.
This paper describes a study designed to determine the possibility of using a dried aerosol solution for calibration in laser ablation inductively coupled plasma mass spectrometry (LA-ICP-MS). The relative sensitivities of tested materials mobilized by laser ablation and by aqueous nebulization were established, and the experimentally determined relative sensitivity factors (RSFs) were used in conjunction with aqueous calibration for the analysis of solid steel samples. To such a purpose a set of CRM carbon steel samples (SS-451/1 to SS-460/1) were sampled into an ICP-MS instrument by solution nebulization using a microconcentric nebulizer with membrane desolvating (D-MCN) and by laser ablation (LA). Both systems were applied with the same ICP-MS operating parameters and the analyte signals were compared. The RSF (desolvated aerosol response/ablated solid response) values were close to 1 for the analytes Cr, Ni, Co, V, and W, about 1.3 for Mo, and 1.7 for As, P, and Mn. Complementary tests were carried out using CRM SS-455/1 as a solid standard for one-point calibration, applying LAMTRACE software for data reduction and quantification. The analytical results are in good agreement with the certified values in all cases, showing that the applicability of dried aerosol solutions is a good alternative calibration system for laser ablation sampling.
本文描述了一项旨在确定使用干燥气溶胶溶液进行激光烧蚀电感耦合等离子体质谱法(LA-ICP-MS)校准的可能性的研究。确定了通过激光烧蚀和雾化所激发的测试材料的相对灵敏度,并将实验测定的相对灵敏度因子(RSF)与水溶液校准结合使用,以分析固体钢样品。为此,使用带膜去溶的微型同心雾化器(D-MCN)通过溶液雾化和激光烧蚀(LA)将一组有证参考物质碳钢样品(SS-451/1至SS-460/1)引入电感耦合等离子体质谱仪。两个系统均采用相同的电感耦合等离子体质谱仪操作参数,并比较分析物信号。对于分析物铬、镍、钴、钒和钨,RSF(去溶气溶胶响应/烧蚀固体响应)值接近1,钼约为1.3,砷、磷和锰为1.7。使用CRM SS-455/1作为单点校准的固体标准进行了补充测试,应用LAMTRACE软件进行数据处理和定量分析。在所有情况下,分析结果与认证值都非常吻合,表明干燥气溶胶溶液的适用性是激光烧蚀采样的一种良好的替代校准系统。