Eliades Theodore, Gioka Christiana, Zinelis Spiros, Eliades George, Makou Margarita
Biomaterials Laboratory, School of Dentistry, University of Athens, Athens, Greece.
World J Orthod. 2004 Spring;5(1):62-6.
To assess the Vickers hardness of plastic brackets, after in vitro aging. The morphology and structure of corresponding retrieved brackets were also studied to assess the reliability of in vitro set-up in imposing effects comparable to intraoral aging.
Four types of plastic brackets were selected. Specimens were immersed in a buffer solution to induce artificial aging; brackets of two groups were retrieved after orthodontic treatment. The wings of brackets were subjected to Knoop hardness testing, and scanning electron microscopy was employed to study the morphology and microstructure of retrieved brackets. The results were statistically analyzed by two-way analysis of variance and Tukey test (alpha = 0.05).
Differences were identified between the brands of brackets, but not between the new and in vitro-aged appliances. The retrieved appliances showed significant reduction in hardness when compared with their as-received counterparts. Evidence of degradation was seen in the retrieved appliances.
The inability of ex vivo experimental configuration to simulate clinical conditions is stressed; several clinical implications related to torque control, friction, and tie-wing strength of brackets are discussed.
评估体外老化后塑料托槽的维氏硬度。还研究了相应回收托槽的形态和结构,以评估体外设置在施加与口腔内老化相当的影响方面的可靠性。
选择四种类型的塑料托槽。将标本浸入缓冲溶液中以诱导人工老化;两组托槽在正畸治疗后回收。对托槽的翼部进行努氏硬度测试,并采用扫描电子显微镜研究回收托槽的形态和微观结构。结果采用双向方差分析和Tukey检验进行统计学分析(α = 0.05)。
托槽品牌之间存在差异,但新的和体外老化器械之间没有差异。与刚收到时相比,回收的器械硬度显著降低。在回收的器械中可见降解迹象。
强调了体外实验配置无法模拟临床情况;讨论了与托槽的扭矩控制、摩擦力和结扎翼强度相关的几个临床意义。