荧光素模式分析是评估用于角膜塑形术的反转几何设计镜片准确性的有效方法吗?

Is fluorescein pattern analysis a valid method of assessing the accuracy of reverse geometry lenses for orthokeratology?

作者信息

Mountford John, Cho Pauline, Chui Wan Sang

机构信息

Suite 7, 12/F T and G Building, 141 Queen Street, Brisbane, QLD 4000, Australia.

出版信息

Clin Exp Optom. 2005 Jan;88(1):33-8. doi: 10.1111/j.1444-0938.2005.tb06661.x.

Abstract

BACKGROUND

The aims of this study were to investigate the relative frequencies of correct identifications of variations in the fit of conventional rigid gas permeable (RGP) lenses and reverse geometry lenses (RGL) from fluorescein pattern analysis by orthokeratology (ortho-k) practitioners and non-ortho-k practitioners and to determine whether fluorescein pattern analysis is sensitive for assessing ortho-k lens fittings.

METHODS

Slides of fluorescein patterns of different lens fittings were shown to the practitioners, who were asked to identify the ideal, flatter, flattest, steeper and steepest lens fittings.

RESULTS

Observed frequencies of correct identifications of most of the conventional RGP lens fittingss were not significantly different from the expected frequencies for both groups of practitioners. The observed frequencies of correct identifications of all of the RGL fittings were either not significantly different or were lower than the expected frequencies.

CONCLUSION

The relative frequencies of correct identifications of fluorescein patterns of both conventional RGP lens and RGL fittings by experienced ortho-k practitioners were not different from those by non-experienced ortho-k practitioners. Practitioners from the two groups were not always able to diagnose conventional RGP lens and RGL fittings adequately from fluorescein pattern analysis alone. Fluorescein pattern analysis alone may not be sufficiently sensitive for assessing ortho-k lens fitting.

摘要

背景

本研究的目的是调查角膜塑形术(ortho-k)从业者和非角膜塑形术从业者通过荧光素图案分析正确识别传统硬性透气性(RGP)镜片和反转几何设计镜片(RGL)适配度变化的相对频率,并确定荧光素图案分析对评估角膜塑形术镜片适配是否敏感。

方法

向从业者展示不同镜片适配的荧光素图案幻灯片,要求他们识别理想、较平、最平、较陡和最陡的镜片适配。

结果

两组从业者对大多数传统RGP镜片适配的正确识别观察频率与预期频率无显著差异。所有RGL适配的正确识别观察频率要么无显著差异,要么低于预期频率。

结论

经验丰富的角膜塑形术从业者对传统RGP镜片和RGL适配的荧光素图案正确识别相对频率与非经验丰富的角膜塑形术从业者无异。两组从业者仅通过荧光素图案分析并不总能充分诊断传统RGP镜片和RGL适配。仅荧光素图案分析对评估角膜塑形术镜片适配可能不够敏感。