Computer-automated tinnitus assessment using patient control of stimulus parameters.

作者信息

Henry James A, Rheinsburg Betsy, Ellingson Roger M

机构信息

VA Medical Center (NCRAR), PO Box 1034, Portland, OR 97207, USA.

出版信息

J Rehabil Res Dev. 2004 Nov-Dec;41(6A):871-88. doi: 10.1682/jrrd.2003.10.0158.

Abstract

A need exists for a standardized tinnitus evaluation tool to measure "acoustic" parameters of tinnitus. An automated tinnitus evaluation system has been developed in this laboratory, consisting of a patient-controlled touch-screen computer monitor, main computer, and Programmable Auditory Laboratory 3000 (PAL 3000) (a custom-built signal conditioning module). The system obtains and records data from tinnitus patients, including hearing thresholds and the pitch and loudness of their tinnitus. New features have been incorporated into the system to make it more user-friendly and practical for clinical application. Using the system, we evaluated 40 individuals with tinnitus to assess within- and between-session reliability of responses. Response reliability has been documented with previous versions of the system. Incorporating the new features has reduced testing time to an average of less than 19 minutes and has resulted in comparable loudness-match reliability and improved pitch-match reliability compared with the previous results. These data support the technique as suitable for clinical application, indicating the need to develop instrumentation for this purpose.