Suppr超能文献

全肘关节置换术因过度使用、C环失效和/或衬套磨损而失败。

Total elbow arthroplasty failure due to overuse, C-ring failure, and/or bushing wear.

作者信息

Wright Thomas W, Hastings Hill

机构信息

Department of Orthopaedic Surgery, University of Florida, Gainesville, FL 32610, USA.

出版信息

J Shoulder Elbow Surg. 2005 Jan-Feb;14(1):65-72. doi: 10.1016/j.jse.2004.04.015.

Abstract

Semiconstrained, linked total elbow arthroplasty gives excellent functional results. However, the complication rate is relatively high, and complications are frequently significant enough to require revision surgery. The purpose of this study is to report on bushing wear and C-ring failure as a mid- to late-term complication of the Coonrad-Morrey total elbow arthroplasty. Two surgeons at different institutions treated 10 patients who had a previous Coonrad-Morrey total elbow arthroplasty and presented with C-ring and/or bushing wear. There were 8 men and 2 women. The mean age at the time of the initial total elbow arthroplasty was 54 years (range, 37-67 years). Time to revision averaged 60 months (range, 9-156 months). Common associated factors were posttraumatic arthritis (3), supracondylar nonunion (5), male sex, young age, and high activity level. A proposed mechanism of failure is postulated.

摘要

半限制型、铰链式全肘关节置换术可取得出色的功能结果。然而,其并发症发生率相对较高,且并发症往往严重到需要进行翻修手术。本研究的目的是报告衬套磨损和C形环失效作为Coonrad - Morrey全肘关节置换术的中晚期并发症情况。来自不同机构的两位外科医生治疗了10例曾接受Coonrad - Morrey全肘关节置换术并出现C形环和/或衬套磨损的患者。其中男性8例,女性2例。初次全肘关节置换术时的平均年龄为54岁(范围37 - 67岁)。翻修的平均时间为60个月(范围9 - 156个月)。常见的相关因素包括创伤后关节炎(3例)、髁上骨不连(5例)、男性、年轻以及活动水平高。本文提出了一种可能的失效机制。