Suppr超能文献

The stability of dry and wet laser imaging systems.

作者信息

AlKhalifah Khaled, Brindhaban Ajit, Al-Ali Hanan, Alhuraibi Anfal

机构信息

Radiologic Sciences Department, Kuwait University Faculty of Allied Health Sciences.

出版信息

Radiol Technol. 2005 Jan-Feb;76(3):192-6.

Abstract

This study investigated the comparative stabilities of wet and dry laser imaging systems using 4 different imagers: 2 wet and 2 dry systems. Fifteen step wedge films, generated by built-in software from each imager, were collected over a period of 25 days. The stability of each system was assessed by computing the average and standard deviation of the percentage coefficients of variation for the optical densities of the steps. Small, but statistically significant, differences between the mean coefficient of variation values were found, with no significant trend toward either system. Wet imaging systems recorded lower speed index values than dry systems, but there was no definite trend toward either type for contrast index. Both imaging systems are equally stable in their performance; however, regular quality control checks are essential to monitor abnormal behavior.

摘要

相似文献

1
The stability of dry and wet laser imaging systems.
Radiol Technol. 2005 Jan-Feb;76(3):192-6.
5
Dry versus conventional laser imagers: film properties and image quality.
Radiology. 1999 Mar;210(3):871-5. doi: 10.1148/radiology.210.3.r99fe03871.
6
Optical density changes in dry-processed films.
Radiol Technol. 2007 Sep-Oct;79(1):9-16.
7
Stability of a dry laser imaging system.
Radiol Technol. 1999 Sep-Oct;71(1):29-34.
8
Image quality of digital chest X-rays: wet versus dry laser printers.
J Digit Imaging. 2001 Sep;14(3):158-62. doi: 10.1007/s10278-001-0015-y.
10
Optical density variations in CT films and their effect on image quality.
Br J Radiol. 2006 May;79(941):425-31. doi: 10.1259/bjr/28579947.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验