用于糖尿病足自旋回波MRI/MRS的双调谐31P/1H鸟笼式射频线圈的射频场均匀性评估。

Evaluation of the RF field uniformity of a double-tuned 31P/1H birdcage RF coil for spin-echo MRI/MRS of the diabetic foot.

作者信息

Greenman Robert L, Rakow-Penner Rebecca

机构信息

Department of Radiology, Beth Israel Deaconess Medical Center and Harvard Medical School, Boston, Massachusetts 02115, USA.

出版信息

J Magn Reson Imaging. 2005 Sep;22(3):427-32. doi: 10.1002/jmri.20372.

Abstract

PURPOSE

To evaluate the B1 field uniformity of a double-tuned birdcage coil designed for (31)P/(1)H MRI/MRS spin-echo (SE) imaging of the metatarsal head region of the foot in neuropathic diabetic patients.

MATERIALS AND METHODS

A low-pass double-tuned (31)P/(1)H RF birdcage coil was constructed to fit over the adult forefoot. Flip angle (FA) maps were created from B1 data acquired at the 3T (31)P (four normal subjects) and (1)H (five normal subjects) frequencies. T2-weighted (T2-W) (1)H images, (31)P rapid acquisition with relaxation enhancement (RARE) images, and composite SE pulse CSI data were acquired to demonstrate the uniformity of the resulting images and data.

RESULTS

The means and standard deviations (SDs) of the range of FAs across the feet of the volunteer subjects indicated good uniformity (the maximum coefficients of variation (CVs) for all of the (31)P and (1)H FA maps were 7.6% and 7.3%, respectively). The FA values across the metatarsal head region indicated a maximum signal intensity variation of +/-3% in a RARE image acquired using an echo train length of 32.

CONCLUSION

A (31)P/(1)H birdcage coil constructed for MRI/MRS studies of the human forefoot provided sufficient signal uniformity of SE data to facilitate accurate (31)P concentration measurements in muscle.

摘要

目的

评估一种为神经性糖尿病患者足部跖骨头区域的(31)P/(1)H磁共振成像/磁共振波谱自旋回波(SE)成像设计的双调谐鸟笼线圈的B1场均匀性。

材料与方法

构建了一个低通双调谐(31)P/(1)H射频鸟笼线圈,以适配成人前足。根据在3T(31)P(四名正常受试者)和(1)H(五名正常受试者)频率下采集的B1数据创建翻转角(FA)图。采集T2加权(T2-W)(1)H图像、(31)P快速采集弛豫增强(RARE)图像以及复合SE脉冲CSI数据,以展示所得图像和数据的均匀性。

结果

志愿者受试者足部FA范围的均值和标准差(SD)表明均匀性良好(所有(31)P和(1)H FA图的最大变异系数(CV)分别为7.6%和7.3%)。在使用回波链长度为32采集的RARE图像中,跖骨头区域的FA值表明最大信号强度变化为±3%。

结论

为人体前足的磁共振成像/磁共振波谱研究构建的(31)P/(1)H鸟笼线圈提供了足够的SE数据信号均匀性,便于准确测量肌肉中的(31)P浓度。