Suppr超能文献

用于儿科胸部的便携式间接平板探测器的实验评估:使用胸部模型在不同曝光条件下与存储磷光体射线照相术的比较。

Experimental evaluation of a portable indirect flat-panel detector for the pediatric chest: comparison with storage phosphor radiography at different exposures by using a chest phantom.

作者信息

Rapp-Bernhardt Ulrike, Bernhardt Thomas M, Lenzen Horst, Esseling Rainer, Roehl Friedrich W, Schiborr Manfred, Theobald-Hormann Iris, Heindel Walter

机构信息

Department of Clinical Radiology, Section of Pediatric Radiology, University of Muenster, 48129 Muenster, Germany.

出版信息

Radiology. 2005 Nov;237(2):485-91. doi: 10.1148/radiol.2372040672. Epub 2005 Sep 16.

Abstract

PURPOSE

To compare the exposure dose requirements and performance of a portable indirect flat-panel detector for pediatric use in the depiction of catheters, simulated pulmonary nodules, and simulated interstitial lung disease with those of storage phosphor radiography.

MATERIALS AND METHODS

Catheters and simulated nodules and subtle interstitial lung disease (miliary, reticular, linear, and ground-glass patterns) were superimposed over an anthropomorphic chest phantom. Images were obtained with different exposures corresponding to simulated speeds of 400 and 800 with a portable flat-panel detector and printed on hard copies. These images were compared with those from storage phosphor radiography at a simulated speed of 400, which is typically used in pediatric radiology. Four independent readers recorded 7200 observations per pattern (for a total of 600 statistically independent observations), and these observations were subjected to receiver operating characteristic (ROC) analysis. Differences were considered significant at a P value of .05.

RESULTS

Catheters over obscured chest areas, nodules 10 mm or smaller and larger than 10 mm over lucent lung, nodules 10 mm or smaller over obscured chest areas, and miliary and linear patterns over lucent lung showed higher areas under the ROC curve (A(z)) with the flat-panel detector at 400 and 800 digital speed compared with storage phosphor radiography. A(z) values for reticular and ground-glass patterns with the flat-panel detector were equal to or less than those with storage phosphor radiography. These differences, however, were not statistically significant.

CONCLUSION

In the detection of catheters, nodules, and almost all interstitial lung disease, A(z) values were higher with the portable flat-panel detector than with storage phosphor radiography at equivalent and reduced speeds. These results suggest that the portable flat-panel detector could be used with reduced exposure dose in pediatric patients.

摘要

目的

比较便携式间接平板探测器在描绘导管、模拟肺结节和模拟间质性肺病时的曝光剂量要求及性能与存储磷光体射线照相术的情况。

材料与方法

将导管、模拟结节和细微的间质性肺病(粟粒状、网状、线状和磨玻璃样)叠加在一个仿真胸部体模上。使用便携式平板探测器以对应模拟速度400和800的不同曝光获取图像,并打印成硬拷贝。将这些图像与存储磷光体射线照相术在模拟速度400(儿科放射学中常用)下获取的图像进行比较。四名独立阅片者针对每种模式记录7200次观察结果(总共600次统计独立观察),并对这些观察结果进行接受者操作特征(ROC)分析。P值为0.05时差异被视为具有统计学意义。

结果

在数字速度400和800下,与存储磷光体射线照相术相比,便携式平板探测器在遮蔽胸部区域的导管、透亮肺内10毫米及以下和大于10毫米的结节、遮蔽胸部区域内10毫米及以下的结节以及透亮肺内的粟粒状和线状模式在ROC曲线下面积(A(z))更高。便携式平板探测器对于网状和磨玻璃样模式的A(z)值等于或低于存储磷光体射线照相术的A(z)值。然而,这些差异无统计学意义。

结论

在检测导管、结节和几乎所有间质性肺病时,便携式平板探测器在等效和降低速度下的A(z)值高于存储磷光体射线照相术。这些结果表明便携式平板探测器可用于儿科患者并降低曝光剂量。