Wu Chung-Yuo, Tsao Jenho, Chou Yi-Hong
Medical Group, Research Center, Micro-Star International Co., Ltd., Taipei City, Taiwan.
Ultrasound Med Biol. 2005 Sep;31(9):1199-210. doi: 10.1016/j.ultrasmedbio.2005.04.015.
The performance of contrast imaging technique is critically influenced by some factors, such as spatial resolution, agent-to-tissue ratio, lifetime of contrast agents and attenuation effect. By using a transducer with higher frequency and higher bandwidth of transmitted signals, the spatial resolution can be improved. Similarly, a better signal-to-interference ratio (SIR) can improve the agent-to-tissue ratio, a lower transmission pressure can prolong the lifetime of contrast agents and a lower center frequency can diminish the attenuation effect. We extend the two-frequency analytic solutions of to approximate the short-pulse responses of microbubbles in a low-amplitude field. Based on their results, there is an expected component near 0 Hz in the spectrum of bubble echoes excited by a short pulse of ultrasound. Here, this component is called the low-frequency response; and it is shown to have a special bandwidth-dependent property and to have potential applications in imaging. We have established the procedure of semi-intermodulated (low-frequency) imaging and the effects of the attenuation effect on imaging resolution, SIR, and signal-to-noise ratio are also analyzed. The obtained experimental images demonstrate that the SIR in semi-intermodulated imaging is better than that in fundamental imaging under various attenuation conditions.
超声造影成像技术的性能受到一些因素的严重影响,如空间分辨率、造影剂与组织的比例、造影剂的寿命以及衰减效应。通过使用具有更高频率和更高发射信号带宽的换能器,可以提高空间分辨率。同样,更好的信干比(SIR)可以提高造影剂与组织的比例,较低的发射压力可以延长造影剂的寿命,而较低的中心频率可以减小衰减效应。我们扩展了[相关内容]的双频解析解,以近似低振幅场中微泡的短脉冲响应。基于这些结果,在短脉冲超声激发的气泡回波频谱中,预计在0 Hz附近有一个分量。在此,该分量被称为低频响应;并且它被证明具有特殊的带宽依赖性特性,并在成像中有潜在应用。我们已经建立了半互调(低频)成像的程序,并分析了衰减效应对成像分辨率、SIR和信噪比的影响。获得的实验图像表明,在各种衰减条件下,半互调成像中的SIR优于基波成像中的SIR。