Hu Zheyi, Gong Qian, Hu Xiaolei, Wang Liqing, Cao Yajie, Cao Wei, Yu Qi, Cheng Zeneng
Department of Biopharmaceutics, School of Pharmaceutical Sciences, Central South University, Changsha, Hunan 410013, PR China.
J Chromatogr B Analyt Technol Biomed Life Sci. 2005 Nov 5;826(1-2):238-43. doi: 10.1016/j.jchromb.2005.09.007. Epub 2005 Sep 23.
This study describes an high-performance liquid chromatographic (HPLC)-UV method for the simultaneous determination of 6beta-hydroxycortisol (6beta-OHF) and cortisol (F) in human urine and plasma. The within-day relative standard deviation of the three concentrations for both analytes was less than 6.2%. Accuracy determined at three concentrations ranged between 95 and 107%. The extraction recoveries were 64.1+/-4.3 and 88.1+/-2.4% at three concentrations for 6beta-OHF and F in urine, respectively. The extraction recoveries were 88.7+/-1.4% at three concentrations for F in plasma. This is the first HPLC method that can simultaneously determine 6beta-OHF and F in human urine and plasma and is suitable for routine assessment of the CYP3A activity expressed as 6beta-hydroxylation clearance.
本研究描述了一种高效液相色谱(HPLC)-紫外检测法,用于同时测定人尿液和血浆中的6β-羟基皮质醇(6β-OHF)和皮质醇(F)。两种分析物在三个浓度水平下的日内相对标准偏差均小于6.2%。在三个浓度水平下测定的准确度在95%至107%之间。尿液中6β-OHF和F在三个浓度水平下的萃取回收率分别为64.1±4.3%和88.1±2.4%。血浆中F在三个浓度水平下的萃取回收率为88.7±1.4%。这是首个能够同时测定人尿液和血浆中6β-OHF和F的HPLC方法,适用于以6β-羟基化清除率表示的CYP3A活性的常规评估。