Sive Barkley C, Zhou Yong, Troop Donald, Wang Yuanli, Little William C, Wingenter Oliver W, Russo Rachel S, Varner Ruth K, Talbot Robert
Climate Change Research Center, Institute for the Study of Earth, Oceans, and Space, University of New Hampshire, Durham, New Hampshire 03824, USA.
Anal Chem. 2005 Nov 1;77(21):6989-98. doi: 10.1021/ac0506231.
An innovative cryogen-free concentrator system for measurement of atmospheric trace gases at the parts per trillion level has been developed with detection by routinely used gas chromatographic methods. The first-generation system was capable of reaching a trapping temperature of -186 degrees C, while the current version can reach -195 degrees C. A Kleemenko cooler is used to create liquid nitrogen equivalent trapping conditions and eliminate the use of solid absorbents, a potential source of artifacts. The method utilizes dual-stage trapping with individual cold regions. The two stages are cooled to -20 and -175 degrees C for water management and sample enrichment, respectively. Both stages house a Silonite-coated stainless steel sample loop; the second stage loop is filled with 1-mm-diameter glass beads, which provide an inert surface area for analyte concentration. In our application, the complete system employed four channels utilizing two flame ionization detectors, one electron capture detector, and a mass spectrometer. The system was automated for unattended operation and was deployed off the New England east coast on Appledore Island to measure a suite of ambient non-methane hydrocarbons, halocarbons, alkyl nitrates, and oxygenated volatile organic compounds during the International Consortium for Atmospheric Research on Transport and Transformation field campaign in summer 2004. This robust system quantified 98 ambient volatile organic compounds with precisions ranging from 0.3 to 15%.
已开发出一种创新的无低温致冷剂浓缩器系统,用于测量万亿分之一水平的大气痕量气体,并通过常规使用的气相色谱方法进行检测。第一代系统能够达到-186℃的捕集温度,而当前版本可达到-195℃。使用克莱门科冷却器来创造液氮等效的捕集条件,并消除使用固体吸收剂,固体吸收剂是潜在的假象来源。该方法利用具有单独冷区的两级捕集。两个阶段分别冷却至-20℃和-175℃,用于水管理和样品富集。两个阶段都装有一个涂有硅酮的不锈钢样品环;第二阶段的环中填充有直径为1毫米的玻璃珠,为分析物浓缩提供惰性表面积。在我们的应用中,完整的系统采用四个通道,利用两个火焰离子化检测器、一个电子捕获检测器和一个质谱仪。该系统实现了自动化无人操作,并部署在新英格兰东海岸的阿普尔多尔岛上,在2004年夏季的国际大气运输与转化研究联盟野外活动期间,用于测量一系列环境非甲烷碳氢化合物、卤代烃、烷基硝酸盐和含氧挥发性有机化合物。这个强大的系统对98种环境挥发性有机化合物进行了定量分析,精度范围为0.3%至15%。