Validity of the GAITRite walkway system for the measurement of averaged and individual step parameters of gait.

作者信息

Webster Kate E, Wittwer Joanne E, Feller Julian A

机构信息

Musculoskeletal Research Centre, La Trobe University, Melbourne, Vic. 3086, Australia.

出版信息

Gait Posture. 2005 Dec;22(4):317-21. doi: 10.1016/j.gaitpost.2004.10.005. Epub 2004 Dec 10.

Abstract

This study compared individual step and averaged spatial and temporal gait parameters measured with an instrumented walkway system (GAITRite) with a three-dimensional motional analysis system (Vicon-512). Ten subjects aged 54-83 years (mean 66.5 years) who had undergone knee replacement surgery participated. Subjects walked across the GAITRite walkway at self-selected comfortable and fast speeds at the same time as the Vicon system recorded the motion of reflective markers attached to the subjects' shoes. Walking speed, cadence, step length and step time variables, averaged across one walk for both systems, showed an excellent level of agreement with intra-class correlation coefficients (ICCs) between 0.92 and 0.99 and repeatability coefficients (RCs) between 1.0% and 5.9% of mean values. Step length and step time variables recorded for each footstep also showed good agreement between the systems at both comfortable and fast speeds (ICCs between 0.91 and 0.99; RCs between 2.6% and 7.8%). Frequency distributions showed that individual step values were within 1.5 cm and 0.02 s on the majority (80-94%) of occasions. These data indicate that the GAITRite system is a valid tool for measuring both averaged and individual step parameters of gait. It is also valid for use in older subjects following knee joint replacement surgery.

摘要

本研究比较了使用仪器化步行道系统(GAITRite)和三维运动分析系统(Vicon - 512)测量的个体步幅以及平均空间和时间步态参数。10名年龄在54 - 83岁(平均66.5岁)且接受过膝关节置换手术的受试者参与了研究。当Vicon系统记录附着在受试者鞋子上的反光标记的运动时,受试者以自我选择的舒适速度和快速速度走过GAITRite步行道。两个系统一次步行的平均步行速度、步频、步长和步时变量显示出极好的一致性水平,组内相关系数(ICC)在0.92至0.99之间,重复性系数(RC)在平均值的1.0%至5.9%之间。在舒适速度和快速速度下,每个脚步记录的步长和步时变量在两个系统之间也显示出良好的一致性(ICC在0.91至0.99之间;RC在2.6%至7.8%之间)。频率分布表明,在大多数(80 - 94%)情况下,个体步幅值在1.5厘米和0.02秒以内。这些数据表明,GAITRite系统是测量步态平均和个体步幅参数的有效工具。它也适用于膝关节置换手术后的老年受试者。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验