Abstract

Primary standardisation was performed on a solution of (18)F using the 4pibeta-gamma coincidence counting efficiency-tracing extrapolation method with (60)Co used as a tracer nuclide. The result was used to calibrate the ANSTO secondary standard ionisation chamber which is used to disseminate Australian activity standards for gamma emitters. Using the secondary activity standard for (18)F, the Capintec CRC-712M dose calibrator at the Australian National Medical Cyclotron (NMC) Positron Emission Tomography (PET) Quality Control (QC) Section was calibrated. The dial setting number recommended by the manufacturer for the measurement of the activity of (18)F is 439. In this work, the dial setting numbers for the activity measurement of the solution of (18)F in Wheaton vials were experimentally determined to be 443+/-12, 446+/-12, 459+/-11, 473+/-15 for 0.1, 1, 4.5 and 9ml solution volumes, respectively. The uncertainties given above are expanded uncertainties (k=2) giving an estimated level of confidence of 95%. The activities determined using the manufacturer recommended setting number 439 are 0.8%, 1.4%, 4.0% and 6.5% higher than the standardised activities, respectively. It is recommended that a single dial setting number of 459 determined for 4.5ml is used for 0.1-9ml solution in Wheaton vials in order to simplify the operation procedure. With this setting the expended uncertainty (k=2) in the activity readout from the Capintec dose calibrator would be less than 6.2%.

摘要

使用以(60)Co作为示踪核素的4πβ-γ符合计数效率追踪外推法,对(18)F溶液进行了初级标准化。该结果用于校准澳大利亚核科学与技术组织(ANSTO)的次级标准电离室,该电离室用于传播澳大利亚γ发射体的活度标准。利用(18)F的次级活度标准,对澳大利亚国立医学回旋加速器(NMC)正电子发射断层扫描(PET)质量控制(QC)部门的Capintec CRC - 712M剂量校准器进行了校准。制造商推荐用于测量(18)F活度的刻度盘设置数为439。在这项工作中,通过实验确定,对于体积分别为0.1、1、4.5和9ml的Wheaton瓶中(18)F溶液活度测量的刻度盘设置数分别为443±12、446±12、459±11、473±15。上述不确定度为扩展不确定度(k = 2),给出的估计置信水平为95%。使用制造商推荐的设置数439确定的活度分别比标准化活度高0.8%、1.4%、4.0%和6.5%。建议为简化操作程序,对于Wheaton瓶中0.1 - 9ml的溶液,使用针对4.5ml确定的单一刻度盘设置数459。采用此设置时,Capintec剂量校准器读出活度的扩展不确定度(k = 2)将小于6.2%。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验