[Comparison of cardiac output measurement in critical patients using transesophageal Doppler or thermodilution: a preliminary report].

作者信息

Andresen Max, Henríquez Mauricio, Mercado Marcelo, Farías Gonzalo, Castillo Carmen, Benítez Carlos, Dougnac Alberto

机构信息

Unidad de Tratamiento Intensivo Médico, Facultad de Medicina, Hospital Clínico Pontificia Universidad Católica de Chile.

出版信息

Rev Med Chil. 2005 Jul;133(7):761-6. doi: 10.4067/s0034-98872005000700002. Epub 2005 Aug 26.

Abstract

BACKGROUND

Cardiac output can be measured non invasively by transesophageal Doppler. This is an alternative to measure it by thermodilution with a catheter in the pulmonary artery.

AIM

To compare both methods of cardiac output measurement.

MATERIAL AND METHODS

Simultaneous measurement of cardiac output by transesophageal Doppler and thermodilution with a catheter in the pulmonary artery in four male critical patients, aged 60+/-12 years, hospitalized in a University Hospital. The Bland and Altman method to compare the concordance between two measurements, was used.

RESULTS

Forty measurements were performed. The results of both methods had a correlation coefficient of 0.98. According to the Bland and Altman method, the difference between both methods was -0.5 L with a precision of 0.52 L/min (95% confidence interval -1.51 to 0.52 L/min). Considering that a change between two sequential measurements is considered significant when the difference is more than 15%, both measurements agreed in 83% of cases, that there was a change in cardiac output.

CONCLUSIONS

Transesophageal Doppler is a promising non invasive technique to measure cardiac output in critical care patients. It becomes a valid alternative to the thermodilution technique. This preliminary experience must be confirmed in a larger series.