Rasch Anton, Dalén Nils, Berg Hans E
Department of Orthopaedic Surgery, Karolinska Institutet at Danderyd Hospital, Stockholm, Sweden.
Arch Phys Med Rehabil. 2005 Dec;86(12):2371-6. doi: 10.1016/j.apmr.2005.05.019.
To evaluate test methods for hip and knee muscle weakness and gait disturbance.
Test-retest.
Orthopedic university clinic.
Ten young (age, 36+/-6 y) and 13 elderly (age, 69+/-8 y) healthy volunteers and 11 patients (age, 69+/-8 y) with unilateral hip osteoarthritis (OA) were tested for muscular strength. Twenty-five volunteers (age, 42+/-14 y) underwent gait analysis.
A dynamometer assessing maximal voluntary isometric force of hip and knee muscles and an optosensor walkway detecting limp were developed. Tests evaluated reproducibility and tolerance in patients with OA and elderly subjects.
Relative coefficient of variation (CV%) and force (in newtons).
CV% for unilateral strength measurements ranged from 7% to 12% for specific muscle groups. CV% for gait parameters ranged from 4% to 8%, except for the double-support phase. Tests were well tolerated, and no patient had to discontinue because of fatigue. Differences related to sex, age, and disease were detected.
Our dynamometer system provides reliable measurements of hip and knee muscle strength in young and old people, and variation is comparable to previous data. Our photocell technique for gait analysis is reliable in people with normal gait. Both methods are attractive because they are affordable, nonstationary, and easy to use.